Electron Microscopy and Elemental Analysis of Defective Drysdale Nucleus Manipulator Instruments Implicated in Surgical Complications - 13/12/22

Doi : 10.1016/j.ajo.2022.09.001 
Lloyd R. Kopecny 1, , Joanna M. Biazik 2, Minas T. Coroneo 3
1 From the School of Clinical Medicine, University of New South Wales Sydney (L.R.K.) 
2 Electron Microscope Unit, Mark Wainwright Analytical Centre, University of New South Wales Sydney (J.M.B.) 
3 and Department of Ophthalmology, Prince of Wales Hospital (M.T.C.), Sydney, NSW, Australia 

Inquiries to Lloyd R. Kopecny, Randwick, New South Wales, AustraliaRandwick, New South WalesAustralia

Highlights

Reusable Drysdale manipulators are widely used in cataract surgery.
Damage or residues can accumulate with instrument reuse.
Scanning electron microscopy found sharp defects in all (5 of 5) reused instruments.
Defects had elemental composition mainly of carbon or aluminum.
Sharp defects may contribute to complications including posterior capsule rupture.

Il testo completo di questo articolo è disponibile in PDF.

Riassunto

PURPOSE

To analyze microtopography of 5 reusable Drysdale nucleus manipulator (DNM) paddled tips for sharp defects and evaluate their elemental composition to determine probable source, investigating 2 instruments (DNM 1 and 4) implicated in causing posterior capsule rupture (PCR) and 3 instruments with sharp edges identified by finger-tip interrogation intraoperatively.

DESIGN

Experimental laboratory investigation.

METHODS

DNM paddled tips were analyzed using scanning electron microscopy (SEM) to evaluate for sharp surface defects (number, dimensions), and subsequently energy dispersive x-ray spectroscopy (EDS) performed on sharp defects to determine their elemental composition.

RESULTS

All reused DNMs analyzed (5 of 5) had significant structural defects on SEM analysis including sharp burrs, cavities and indentations, surface debris or residues, and roughening, compared to the new instrument (DNM 3, control) which had no defects. DNM 1 had 2 sharp defects, a larger 14 × 76-µm one and a craterlike 167 × 220-µm defect containing debris. EDS found that DNM 2 had 3 of 4 burrs composed mainly of carbon, the fourth of mixed composition (calcium, sulfur, oxygen); DNM 4 had 1 small burr, EDS significant for carbon; DNM 5 had 3 prominent burrs, the largest of 20 × 28 µm, 2 composed of aluminum, and some carbon residue. DNM 6 had 1 burr composed of aluminum and 3 prominent cavity defects, the largest covering 781 µm2.

CONCLUSION

Reusable DNMs are widely used in cataract surgery. Sharp carbon- or aluminum-containing burrs were detected on all reused instruments analyzed together with 1 burr of mixed composition, originating from (1) organic residues, (2) instrument fragments, or (3) salt and contaminant deposits. Sharp defects may contribute to capsular damage including PCR, and residues may pose other safety concerns. Therefore, we support development of a quality, reliable single-use alternative instrument and further encourage careful inspection of all reusable instruments principally by finger-tip interrogation for sharp edges preuse.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


 Supplemental Material available at AJO.com.


© 2022  Elsevier Inc. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 245

P. 102-114 - gennaio 2023 Ritorno al numero
Articolo precedente Articolo precedente
  • Full-Thickness Macular Hole: Are Supra-RPE Granular Deposits Remnants of Photoreceptors Outer Segments? Clinical Implications
  • Andrea Govetto, Daniela Bacherini, Mario R. Romano, Martina Matteucci, Sara Lucchini, Lorenzo Cifarelli, Francesco Dragotto, Matteo Cereda, Salvatore Parrulli, Giovanni Staurenghi, Fabrizio Giansanti, Gianni Virgili, Ferdinando Bottoni, Paolo Radice
| Articolo seguente Articolo seguente
  • Large Internal Limiting Membrane Tears: Prevalence, Clinical Characteristics, and Surgical Utility
  • Asad F. Durrani, Robert A. Hyde, Mark W. Johnson

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.