Validation of a quantitative screening method by liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS) from a dried blood spot (DBS) - 03/03/25

Doi : 10.1016/j.toxac.2025.01.022 
Robin Tiphaine , Hubert Chloé, Larabi Islam Amine, Alvarez Jean-Claude
 Hauts-de-seine, Laboratoire de pharmacologie-toxicologie, CHU Raymond Poincaré (AP–HP), Garches, France 

Corresponding author.

Résumé

Aim

To develop and validate a quantitative screening method for a panel of 300 compounds by LC-MS/MS using a DBS.

Method

A drop of blood (10μL) is applied to HemaXis DB 10 filter paper (HMX-DB-10) and left to dry for 3hours. The compounds of interest are then extracted by a two-step liquid-liquid extraction: first with 2mL of a hexane/ethyl acetate mixture (50/50) in an alkaline environment (pH 9.7), followed by 2mL of a chloroform/isopropanol mixture (80/20). After evaporation of the supernatants, the residue is reconstituted with 80μL of a mixture of 2mM ammonium formate (0.1% formic acid) and acetonitrile (50/50). In all, 10μL is then injected into the system. Separation is performed on a HypersilGold PFP column (100×2.1mm, 1.9μm) using a gradient mode with 2mM formate buffer (0.1% formic acid) and acetonitrile. Detection is performed in MRM mode using a triple quadrupole mass spectrometer (TSQ Altis, ThermoFisher®).

Results

The method was validated for several panels of compounds, including traditional drugs of abuse (cocaine and its metabolites, 13 amphetamines, 28 opioids), NPS (12 synthetic benzodiazepines, 29 cathinones, 26 synthetic cannabinoids, 26 new synthetic opioids, other NPS), as well as several drug classes (30 benzodiazepines, 37 antipsychotics, 30 antidepressants, 12 antiepileptics, 18 β-blockers, 8 anesthetics, 8 sartans, etc.). The limits of quantification (LOQ) ranged from 5ng/mL (86% of compounds) to 10ng/mL (14% of compounds), with bias and CV ≤20%. Linearity was validated over a range of 5–10 to 500ng/mL with satisfactory performance in terms of accuracy and precision (bias and CV ≤15%). Analysis of a 200ng/mL quality control showed accuracy (bias+precision ≤15%). No carry-over effect was observed in blanks injected after the high point (500ng/mL). Selectivity was verified using negative blood samples, with no significant interferences. Patient samples were quantified, and results were compared to those obtained by the routine blood screening LC-MS/MS method used in the laboratory. The concentrations obtained were similar between the two techniques. Studies on stability, recovery, and reproducibility are on going.

Conclusion

The use of DBS for toxicological analyses or pharmacological therapeutic drug monitoring (TDM) is gaining popularity. This is due to several advantages: a simplified sampling procedure, a small sample volume, and good analyte stability on this type of substrate.

The adoption of DBS could also improve patient compliance, thanks to the simplicity of the collection (which can be done by the patient themselves), easy transport (by post), and storage without constraints (at room temperature). Additionally, the small volume of blood required could be a solution in specific contexts, such as pediatric sampling or multiple analyses, where samples are often limited.

However, the analysis of very small blood volumes (10μL) poses a significant challenge, requiring the use of instruments with very high sensitivity. Another important constraint is the nature of the matrix being analyzed: whole blood. Since TDM is primarily performed on plasma, models for predicting plasma concentrations from DBS results need to be developed and validated. Blood viscosity also affects the results, as it can impact sample homogeneity or extraction efficiency. Hematocrit determines blood viscosity, and its effect should be evaluated during analytical validation.

Advances in analytical techniques enable increasingly higher sensitivity. The results demonstrate the potential of DBS for therapeutic drug monitoring applications or forensic toxicology.

Le texte complet de cet article est disponible en PDF.

Plan


© 2025  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 37 - N° 1S

P. S19 - mars 2025 Retour au numéro
Article précédent Article précédent
  • How to respond to real time drugs-alerts? Comparison between French and Belgian early warning systems
  • Sabrina Cherki, Margot Balcaen
| Article suivant Article suivant
  • Analysis of oral fluid samples results by LC-MS/MS and comparison with law enforcement screening: A 6-year study
  • Charlotte Mayer, Adeline Knapp-Gisclon, Jean-Claude Alvarez

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.