S'abonner

Does operator experience affect the accuracy of guided palatal miniscrew insertion via surgical guide? An in-vitro study - 19/03/25

Doi : 10.1016/j.ortho.2025.100995 
Amedeo Salomone 1, Mario Palone 1, , Francesca Cremonini 1, Giuliano Maino 1, Emanuele Paoletto 2, Marta Cappelletti 1, Luca Lombardo 1
1 Postgraduate School of Orthodontics, University of Ferrara, Via Luigi Borsari 46, Ferrara 44121, Italy 
2 Lab Orthomodul, Thiene 36016, Italy 

Mario Palone, Postgraduate School of Orthodontics, University of Ferrara, Via Luigi Borsari 46, Ferrara 44121, Italy.Postgraduate School of Orthodontics, University of FerraraVia Luigi Borsari 46Ferrara44121Italy

Summary

Introduction

To compare the in-vitro accuracy of guided palatal miniscrew insertion comparing expert and inexpert clinicians.

Material and methods

Twenty-one synthetic bone models, derived from a single master model, were acquired to simulate the clinical act of miniscrew insertion. Digital planning and CAD/CAM surgical guide manufacturing were executed by matching the CBCT of the master model with its corresponding STL file. The insertion of two palatal miniscrews in the anterior paramedian region was planned. The operators (mean age 35 years±5 years; 11 males and 9 females) were divided into two sub-groups (inexperienced and experienced), and the miniscrews inserted using a standardized procedure. Linear and angular discrepancies between planned and inserted miniscrew positions were then evaluated at the level of head and tip point by superimposing the reference model (derived from digital planning) with the 20 working models (derived from scanning after miniscrew insertion). Absolute accuracy and comparison between the sub-groups were assessed using a one-sample Wilcoxon test (P<0.05).

Results

Regardless of experience, a statistically significant difference in all investigated measurements was found. However, no statistically significant differences were detected between the two sub-groups, except for the sagittal discrepancy at the head, with the inexperienced group being less accurate (P=0.002).

Conclusions

The use of a CAD/CAM surgical guide ensures comparable accuracy between inexperienced and experienced clinicians, excepting some outlier discrepancies among the inexpert subjects. Although there are differences in accuracy between the planned and achieved miniscrew position, these differences do not appear to be clinically significant.

Le texte complet de cet article est disponible en PDF.

Keywords : Palate, Miniscrew, Accuracy, CAD/CAM surgical guide


Plan


© 2025  CEO. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 23 - N° 3

Article 100995- septembre 2025 Retour au numéro
Article précédent Article précédent
  • Comparative evaluation of novel framework-supported 3-dimensional facial scanning using smartphone device with freehand facial scanning in patients seeking orthodontic treatment: A cross-sectional study
  • Aditya Talwar, Isha Duggal, Harpreet Grewal, Neha Bhutiani, Navneet Singh, Saurabh Kaushik
| Article suivant Article suivant
  • Comparison of stress distribution and displacement pattern of maxillary expansion in craniomaxillary complex using rapid maxillary expander and maxillary skeletal expander: A finite element model analysis
  • Monalisa Goswami, Anshu Sahu, Raghu Ranjan Prasad, Parul Priya, Parul Priya, Shovan Roy, Abhijith A

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.