Contrasting Heating Patterns and Efficiency of the Prostatron and Targis Microwave Antennae for Thermal Treatment of Benign Prostatic Hyperplasia - 08/09/11

Doi : 10.1016/S0090-4295(98)00142-3 
Thayne R. Larson a, , Michael L. Blute b, Jeffrey L. Tri c, Sidney V. Whitlock c
a Department of Urology, Mayo Clinic, Scottsdale, Arizona, USA 
b Department of Urology, Mayo Clinic, Rochester, Minnesota, USA 
c Division of Engineering and Technical Services, Mayo Clinic, Rochester, Minnesota, USA 

*Reprint requests: Thayne R. Larson, M.D., Department of Urology, 13400 East Shea Boulevard, Scottsdale, AZ 85259

Abstract

Objectives.

To determine the design and performance characteristics of two microwave antennae for use in thermal treatment of benign prostatic hyperplasia.

Methods.

Prostatron and Targis antennae were subjected to detailed physical examination and measurement. The heating patterns generated by these two types of antennae were characterized in detail using tissue-equivalent phantoms. Measurements of return loss as a function of frequency were conducted to evaluate the capacity of the antennae for impedance matching. Percent reflected power was calculated from the return loss results to provide a relative measure of potential for efficient delivery of thermal energy.

Results.

The Prostatron antenna was found to be a monopole design consisting of a coaxial cable with a 3.3- cm length of inner conductor exposed at the tip. The Targis antenna was observed to be a dipole design with a 2.8-cm helical coil attached through a ground connection and a tap point to a coaxial cable. The heating pattern of the Targis antenna was symmetric; that of the Prostatron was asymmetric with substantial back heating along the catheter axis in the direction of the microwave power source. The mean extension of the 30°C isotherm in the direction of the power source with the Prostatron antenna (71.5 mm; 95% confidence interval [CI], 63.4 to 79.6 mm) was 55% greater (P <0.0005) than that with the Targis antenna (46.0 mm; 95% CI, 38.2 to 53.8 mm). Return loss with the Targis antenna declined sharply to a relative minimum value of −32.9 dB (95% CI, −73.8 to 8.0 dB) at 915 MHz, providing evidence of this antenna’s capacity for impedance matching; little change was observed with the Prostatron in return loss over a frequency range 100 MHz above and below this antenna’s standard operating frequency of 1296 MHz. The mean reflected power of the Targis antenna (0.4%; 95% CI, 0.0% to 1.4%) was lower by more than 20-fold (P = 0.036) than that of the Prostatron antenna (11.0%; 95% CI, 3.4% to 18.7%); thus, the potential for efficient operation was greater with the Targis than the Prostatron antenna.

Conclusions.

The Targis microwave antenna was found to provide a more targeted heating pattern and have a capacity for more efficient thermal energy delivery than the Prostatron antenna. These differences observed in vitro could potentially translate into clinical advantages in vivo, such as improved tolerability of microwave treatment, reduced risk of complications, greater thermoablative efficacy, and scalability.

Le texte complet de cet article est disponible en PDF.

Plan


© 1998  Elsevier Science Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 51 - N° 6

P. 908-915 - juin 1998 Retour au numéro
Article précédent Article précédent
  • Long-Term Evaluation of Tamsulosin in Benign Prostatic Hyperplasia: Placebo-Controlled, Double-Blind Extension of Phase III Trial
  • Herbert Lepor, for the Tamsulosin Investigator Group 1A list of the Tamsulosin Investigator Group appears in the . Appendix 1
| Article suivant Article suivant
  • Laparoscopy for the Acute Abdomen in the Postoperative Urologic Patient
  • John J. Bauer, Peter G. Schulam, Howard S. Kaufman, Robert G. Moore, Pierce B. Irby, Louis R. Kavoussi

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.