Matrix using SPE and LC-MS/MS - 07/05/15

Doi : 10.1016/j.toxac.2015.03.084 
A. Leffler , S. Huq, S. Sadjadi, S. Orlowicz
 Phenomenex Inc., Torrance, CA, USA 

Corresponding author.

Résumé

Background

Drug analysis from whole blood is gaining popularity due to a more complete measurement of analytes in the biological system. Historically, most drug screens have been performed using immunoassay and similar technologies. These techniques often lack specificity and sensitivity necessary for today's complex legal requirements. In this poster, we demonstrate a fast, robust technique for the screening of a broad range of drugs from whole blood utilizing SPE and tandem mass spectrometry. Forty-one compounds, which include opiates (both natural and synthetic), amphetamines, benzodiazepines, phenylpiperidines, a muscle relaxant, and some illicit drugs (PCP and cocaine metabolite) were tested.

Objective

Utilizing advanced sample preparation techniques we simplify a complicated matrix to allow for a fast and successful multi-component analysis by LC-MS/MS.

Method

Six pretreatment options for whole blood were investigated. The volume of whole blood used was 0.25mL. A 5% zinc sulfate and acetonitrile combination worked for most classes. Benzodiazepines had a better overall response with 10% MeOH in ACN. However, there was not a single pretreatment procedure that worked well for all classes of compounds. The supernatant was loaded onto the Strata-XC 30mg/3mL cartridges. Salts and most endogenous components were removed using a two-step, 0.1% formic acid and 30% methanol, wash. Analytes were eluted using 2×500 uL of ammoniated IPA/ethyl acetate solution. Samples were then acidified and evaporated to dryness at 40–45°C under a gentle nitrogen stream. The dry residue was re-suspended in mobile phase and injected onto the column. The chromatographic separation was performed on a Kinetex 2.6um Biphenyl 50×3.0mm column. The best choice of mobile phase was 0.1% formic acid in water and 0.1% formic acid in methanol. The detection was achieved on an AB Sciex Triple Quad 4500 and/or a 4000 QTrap LC-MS/MS system. All analytes were detected under positive polarity and MRM scan function using two mass transitions. A MRM-IDA-EPI combination scan was used on a small sample set to verify the analyte purity.

Results

A detection limit of 10ng/mL was achieved for all tested analytes using this method. Tramadol, propoxyphene, fentanyl, carisoprodol and meprobamate produced strong signals. The absolute lowest detection levels for these analytes have not yet been determined. Although a full calibration method was not vigorously tested, the upper end of the calibration range was deemed to be 1000ng/mL. A quadratic fit with 1/× weighting factor was employed for calibration purposes. Early eluting analytes were well separated from the ion suppression zone by a retention factor of two, or by three times the dead volume. Isomeric/isobaric compounds were completely resolved by the biphenyl LC column by at least a factor of 2. A resolution value of 3.2 was obtained for morphine and hydromorphone.

Conclusion

The described procedure provides a simple and reliable solution to screen a wide range of illicit and pain control compounds from a complicated matrix.

Le texte complet de cet article est disponible en PDF.

Plan


© 2015  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 27 - N° 2S

P. S55 - juin 2015 Retour au numéro
Article précédent Article précédent
  • Development and validation of an analytical method for quantification of 15 non-tricyclic antidepressants in serum with UPLC-MS/MS
  • S. Resende, C. Deschrijver, E. Van De Velde, A. Verstraete
| Article suivant Article suivant
  • Identification de cyclobenzaprine lors de recherche de toxiques : impureté ou réelle consommation ?
  • J. Bois-Maublanc, T. Francia, L. Got, S. Lefeuvre

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Elsevier s'engage à rendre ses eBooks accessibles et à se conformer aux lois applicables. Compte tenu de notre vaste bibliothèque de titres, il existe des cas où rendre un livre électronique entièrement accessible présente des défis uniques et l'inclusion de fonctionnalités complètes pourrait transformer sa nature au point de ne plus servir son objectif principal ou d'entraîner un fardeau disproportionné pour l'éditeur. Par conséquent, l'accessibilité de cet eBook peut être limitée. Voir plus

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2026 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.