Primary cup stability in THA with augmentation of acetabular defect. A comparison of healthy and osteoporotic bone - 06/10/15

Doi : 10.1016/j.otsr.2015.07.007 
F. Amirouche a, , G. Solitro a, S. Broviak a, W. Goldstein a, b, M. Gonzalez a, R. Barmada a
a Department of orthopaedics, university of Illinois at Chicago, 835, S. Wolcott avenue, 60616 Chicago, Illinois, USA 
b Illinois bone and joint institute, 9000, Waukegan road, 60053 Morton Grove, Illinois, USA 

Corresponding author. Department of orthopaedics, university of Illinois at Chicago, 835, S. Wolcott avenue, room E270, 60612 Chicago, IL, USA. Tel.: +312 996 7161; fax: +312 996 9025.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Background context

Reconstruction of acetabular defect has been advocated as standard procedure in total hip arthroplasty. The presence of bony defects at the acetabulum is viewed as a cause of instability and acetabular wall augmentation is often used without proper consideration of surrounding bone density. The initial cup-bone stability is, however, a challenge and a number of studies supported by clinical follow-ups of patients suggested that if the structural graft needs supporting more than 50% of the acetabular component, a reconstruction cage device spanning ilium to ischium should be preferred to protect the graft and provide structural stability. This study aims to (1) investigate the relationship between cup motion and bone density and (2) quantify the re-distribution of stress at the defect site after augmentation.

Hyphotesis

Paprosky type I or II, acetabular defects, when reconstructed with bone screws supported by bioabsorbable calcified triglyceride bone cement are significantly less effective for osteoporotic bone than healthy bone.

Materials and methods

Acetabular wall defects were reconstructed on six cadaveric subjects with bioabsorbable calcified triglyceride bone cement using a re-bar technique. Data of the specimen with higher bone density was used to validate a Finite Element Model. Values of bone apparent density ranging from healthy to osteoporotic were simulated to evaluate (1) the cup motion, through both displacement and rotation, (2) and the von Mises stress distribution.

Results

Defect reconstruction with bone screws and bioabsorbable calcified triglyceride bone cement results in a re-distribution of stress at the defect site. For a reduction of 65% in bone density, the cup displacement was similar to a healthy bone for loads not exceeding 300N, as load progressed up to 1500N, the reconstructed defect showed increase of 99μm (128%) in displacement and of 0.08° in rotation angle.

Conclusions

Based on the results, we suggest that an alternative solution to wall defect augmentation with bone screws supported by bioabsorbable calcified triglyceride bone cement, be used for osteoporotic bone.

Level of evidence

Level IV, experimental and cadaveric study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Arthroplasty, Replacement, Hip prosthesis, Bone cements, Osteoporosis


Mappa


 This work was partially supported by Aurelio M. Caccomo family foundation.


© 2015  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 101 - N° 6

P. 667-673 - ottobre 2015 Ritorno al numero
Articolo precedente Articolo precedente
  • Excellent short-term results of hip resurfacing in a selected population of young patients
  • A. Lons, A. Arnould, T. Pommepuy, E. Drumez, J. Girard
| Articolo seguente Articolo seguente
  • Subtrochanteric femoral fractures treated with the Long Gamma3® nail: A historical control case study versus Long trochanteric Gamma nail®
  • D. Georgiannos, V. Lampridis, I. Bisbinas

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.