Right Parasternal Electrode Configuration Converts a Failed Electrocardiographic Screening to a Pass for Subcutaneous Implantable Cardioverter-Defibrillator Implantation - 18/11/15

Doi : 10.1016/j.hlc.2015.07.015 
N.Y. Chan, MBBS FRCP FACC FHRS , H.C. Yuen, MBBS, N.S. Mok, MBBS FRCP
 Department of Medicine & Geriatrics, Princess Margaret Hospital, Hong Kong 

Corresponding author at: Room 223, Block J, Princess Margaret Hospital, 2-10 Princess Margaret Hospital Road, Lai Chi Kok, Kowloon, Hong Kong. Tel.: +(852) 29903725; fax: +(852) 29903329.

Riassunto

Background

Pre-operative electrocardiographic (ECG) screening before subcutaneous implantable cardioverter-defibrillator (SICD) implantation is essential to prevent T-wave oversensing and inappropriate shocks. The failure rate of ECG screening was reported to be up to 8% when only two body positions were tested.

Method

Three subcutaneous ECG vectors represented by lead I, II and III were obtained in standing, supine, sitting and squatting positions. A patient qualified if the ECG in any same lead passed in all four positions. We report a 31-year-old man with idiopathic ventricular fibrillation who failed ECG screening for SICD implantation with the conventional left parasternal electrode (LPS) configuration in all three subcutaneous ECG vectors. Right parasternal electrode (RPS) configuration with left arm and right arm ECG electrodes positioned 1cm lateral to right sternal border was attempted for screening.

Result

The amplitude of the QRS complex was significantly larger in the RPS compared to the LPS configuration in lead III and the patient passed the ECG screening in four body positions. He underwent successful SICD implantation with RPS approach with appropriate sensing both during the procedure and exercise treadmill test four weeks later. Ventricular fibrillation was successfully converted with 65J standard polarity shock during the procedure and no ICD shock was experienced by the patient on six-month follow-up.

Conclusion

RPS configuration may be considered in patients who fail the ECG screening with the conventional LPS approach for SICD implantation.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Subcutaneous implantable cardioverter-defibrillator, Screening, Sudden cardiac death, Right parasternal, Subcutaneous electrocardiography



© 2015  Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 24 - N° 12

P. e203-e205 - dicembre 2015 Ritorno al numero
Articolo precedente Articolo precedente
  • Response to ‘Ex-vivo Donor Heart Perfusion: Testing the Limits of Cardiac Resilience’
  • N.L. Stamp, R.I. Larbalestier
| Articolo seguente Articolo seguente
  • Retrograde Type A Aortic Dissection Treated with Continuous Perfusion “Branch-First” Aortic Arch Replacement Technique
  • Louise C. Robertson, Damien Holdaway, Cheng-Hon Yap

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.