Dose reduction with iterative reconstruction in multi-detector CT: What is the impact on deformation of circular structures in phantom study? - 10/02/16

Doi : 10.1016/j.diii.2015.06.019 
J. Greffier a, , F. Macri a, A. Larbi a, A. Fernandez a, F. Pereira a, C. Mekkaoui a, b, 1, J.-P. Beregi a, 1
a EA 2415, Department of Radiology, Nîmes University Hospital, boulevard Professeur-Robert-Debré, 30029 Nîmes cedex, France 
b Martinos Centre for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, United States of America 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Objectives

To evaluate the distortion of circular structures induced by the increased image noise related to dose reduction and to assess the effect of iterative reconstruction (IR).

Methods

MDCT acquisitions were performed with 120 kVp for 200/100/60/40/20mAs with 100%/50%/30%/20%/10% of dose. Raw data were reconstructed by filtered back projection (FBP) and with two IR strengths. Image quality indices referred to water and acrylic were measured on a quality image phantom. Areas, perimeters, circularity were measured on the circular inserts of 4.8, 7.9 and 11.1mm on a morphological phantom.

Results

Dose reduction resulted in increased image noise and in decreased signal to noise ratio and contrast to noise ratio. IR improved these indices for the same dose without affecting the signal (number CT) and spatial resolution (modulation transfer function). The values of area, perimeter and circularity were altered compared to the actual value and the inserts were visually deformed with the dose reduction. IR improved these three parameters. Image quality indices, areas, perimeters and circularity of inserts were similar between the acquisition at 100% of the dose in FBP, 50% in strength-3 and 30% in strength-5 with different curves of noise power spectrum.

Conclusion

IR associated to 70% of dose reduction modifies the images smooth (NPS) but maintains adequate image quality indices without causing distortions of circular structures.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Computed tomography, Iterative reconstruction, Image quality, Deformation of circular structures, Noise power spectrum

Abbreviations : CNR, CTDI, FBP, FOV, IR, MDCT, MTF, NPS, PSF, ROI, SAFIRE, SNR, TTF


Mappa


© 2015  Éditions françaises de radiologie. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 97 - N° 2

P. 187-196 - febbraio 2016 Ritorno al numero
Articolo precedente Articolo precedente
  • Efficacy of diffusion-weighted magnetic resonance imaging in the diagnosis and staging of endometrial tumors
  • I.S. Kececi, M.S. Nural, K. Aslan, M. Danac?, M. Kefeli, M. Tosun
| Articolo seguente Articolo seguente
  • The short- and long-term effectiveness of transcatheter arterial embolization in patients with intractable hematuria
  • M. Korkmaz, B. ?anal, B. Aras, H. Bozkaya, C. Ç?nar, S. Güneyli, M. Gök, G. Adam, F. Düzgün, ?. Oran

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.