Radiation dose and diagnostic accuracy of high-pitch dual-source coronary angiography in the evaluation of coronary artery stenoses - 31/03/16

Doi : 10.1016/j.diii.2015.10.008 
M. Koplay a, , H. Erdogan a, A. Avci b, M. Sivri a, K. Demir b, I. Guler a, L.S. Demir c, Y. Paksoy a
a Selcuk University, The Central Campus, Medical Faculty, Department of Radiology, 42075 Konya, Turkey 
b Selcuk University, Medical Faculty, Department of Cardiology, 42075 Konya, Turkey 
c Necmettin Erbakan University, Meram Medical Faculty, Department of Public Health and Medical Statistics, 42060 Konya, Turkey 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Purpose

“Flash Spiral” imaging is a new prospective ECG-triggered spiral scan mode that uses a very high-pitch for coronary computed tomography angiography (CTA). This enables complete image acquisition within one cardiac cycle with a very low radiation exposure. The aim of this study was to investigate the diagnostic accuracy, image quality, and effective radiation dose of prospectively ECG-triggered high-pitch spiral method (Flash spiral mode) of coronary CTA using dual-source technology for the evaluation of coronary artery stenoses.

Material and methods

The study included 186 consecutive patients (115men, 71women; mean age: 53.37 years) who underwent coronary CTA. Coronary CTA was performed with a 128×2-slice dual-source CT (Somatom Definition Flash, Siemens, Germany) using a prospectively ECG-triggered high-pitch spiral mode. Patients were divided into three groups according to heart rate (≤65bpm, 66–75bpm, ≥76bpm) and body mass index (BMI) (20–24kg/m2, 25–29kg/m2, 30–34kg/m2) values. The correlation between heart rates, image quality and BMI values are investigated. A four-point scale (1=excellent, 4=poor/non-diagnostic) was used to rank the comparative image quality. Effective radiation doses were calculated. Also the correlation between radiation dose, sex and BMI values were investigated. In addition, diagnostic accuracy of CTA for detection of significant (≥50%) coronary artery stenoses was compared with invasive coronary angiography findings of 612vessel segments in 38patients. The sensitivity, specificity, positive predictive value, negative predictive value and diagnostic accuracy rate were calculated.

Results

A total of 2976coronary artery segments were present. Of all coronary artery segments, 2381 (80%) had an image quality score of 1, and 1.5% segments were rated as “poor/non-diagnostic”. When the correlation between heart rate and image quality is investigated, there was a significant difference between ≤65bpm and ≥76bpm groups. However, there was no significant difference between ≤65bpm and 66–75bpm groups. The mean effective dose was found as 1.3mSv (min: 0.5, max: 2.4mSv). The correlation between effective radiation dose and BMI was moderate. However, the correlation between gender and radiation dose was significant. Sensitivity, specificity, negative predictive value and accuracy of coronary CTA on a per-vessel segment were 90.1%, 97.4%, 98.6%, and 96.5%, respectively.

Conclusions

Prospectively ECG-triggered high-pitch spiral mode coronary CTA provides high image quality and diagnostic accuracy, with very low radiation dose for evaluation and exclusion of coronary artery stenoses.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Coronary angiography, Dual source CT, High-pitch spiral acquisition, Image quality, Radiation dose


Mappa


© 2015  Éditions françaises de radiologie. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 97 - N° 4

P. 461-469 - Aprile 2016 Ritorno al numero
Articolo precedente Articolo precedente
  • Comparison between Doppler ultrasonography and multiphase multidetector-row computed tomography in the detection of liver involvement in Rendu-Osler disease: An analysis of 62 patients
  • A. Cavel, A. Bleuzen, P. Bertrand, F. Patat, J.-P. Cottier
| Articolo seguente Articolo seguente
  • Peripheral hypertrophic neuropathy due to leprosy: Ultrasound and MR imaging findings
  • P. Pottecher, B. Flageul, E. Sibileau, J.-D. Laredo, V. Bousson

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2024 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.