Subacromial temperature profile during bipolar radiofrequency use in shoulder arthroscopy. Comparison of Coblation® vs. VAPR® - 29/11/17

Doi : 10.1016/j.otsr.2017.02.011 
V. Huynh , O. Barbier, X. Bajard, A. Bouchard, D. Ollat, G. Versier
 Service de chirurgie orthopédique et traumatologique, hôpital d’instruction des armées Bégin, 69, avenue de Paris, 94160 Saint-Mandé, France 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Background

The use of bipolar electrodes for arthroscopic procedures carries a theoretical ex vivo risk of inducing burn injuries. Few studies have measured the in vivo temperatures produced by bipolar electrodes during arthroscopy, and their results are conflicting. The objective of this study was to evaluate the temperature profile within the subacromial space during shoulder arthroscopy with two different electrode systems.

Hypothesis

The primary hypothesis was that the two electrode systems produced similar temperature variations and peak temperatures. The secondary hypothesis was that neither electrode system produced irrigation-fluid temperatures above the tissue-damage threshold.

Material and methods

A comparative, prospective, single-centre, single-surgeon, single-blind study was conducted to compare the Coblation® system (Smith&Nephew, Andover, MA, USA) and the VAPR® system (DePuy Synthes Mitek Sports Medicine, Raynham, MA, USA) in 13 patients undergoing shoulder arthroscopy. A temperature probe inserted into the subacromial space was used to record temperatures at 10-second intervals for 60seconds during continuous radiofrequency application.

Results

Mean baseline temperature was 21.4±0.7°C with VAPR® and 23.0±2.2°C with Coblation®. No significant between-group differences were found during the first 40seconds. The mean peak temperature reached after 60seconds was 25.0±1.9°C with VAPR® and 27.9±2.8°C with Coblation® (P<0.05).

Discussion

Few studies have compared the in vivo temperatures produced during arthroscopy by different electrode systems. In vivo studies have established that temperature increases can cause tissue damage, particularly to chondrocytes, and that the irrigation flow rate plays a key role in lowering the in vivo temperatures. Our study showed a significant difference between the two electrode systems after 50seconds of use, with lower temperatures with the VAPR®. Nevertheless, neither system increased the irrigation-fluid temperatures above the tissue-damage threshold. Both systems can be used safely, provided the manufacturer instructions are followed and the irrigation system is effective.

Level of evidence

II (prospective randomized trial).

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Temperature, Coblation, VAPR, Shoulder, Arthroscopy


Mappa


© 2017  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 103 - N° 4

P. 489-491 - giugno 2017 Ritorno al numero
Articolo precedente Articolo precedente
  • Learning the skills needed to perform shoulder arthroscopy by simulation
  • P.-S. Marcheix, G. Vergnenegre, F. Dalmay, C. Mabit, J.-L. Charissoux
| Articolo seguente Articolo seguente
  • Changes of the bone structure after cap resurfacing arthroplasty of the humeral head
  • L.V. von Engelhardt, M. Hahn, T. Schulz, K. Peikenkamp, J. Jerosch

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.