Shoulder patient-specific guide: First experience in 10 patients indicates room for improvement - 23/02/18

Doi : 10.1016/j.otsr.2017.11.005 
J. Berhouet a, b, , M. Rol a, C. Spiry a, M. Slimane b, C. Chevalier a, L. Favard a
a Service d’orthopédie traumatologie, faculté de médecine de Tours, université François-Rabelais-de-Tours, CHRU Trousseau, 1C, avenue de la République, 37170 Chambray-Les-Tours, France 
b Équipe reconnaissance de forme et analyse de l’image, laboratoire d’informatique EA6300, école d’ingénieurs polytechnique universitaire de Tours, université François-Rabelais-de-Tours, 64, avenue Portalis, 37200 Tours, France 

Corresponding author at: Service d’orthopédie traumatologie 2A, université François-Rabelais-de-Tours, CHRU Trousseau, avenue de la République, 37170 Chambray-Les-Tours, France.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Background

Implantation of the glenoid component of a total shoulder prosthesis can be facilitated by using a patient-specific guide (PSG) designed to ensure replication of the preoperatively planned position. The objective of this study was to assess the reliability and accuracy of a PSG in replicating the planned glenoid component position during total shoulder arthroplasty (TSA).

Hypothesis

Additional criteria should be used for 3D preoperative planning and PSG design to further improve the accuracy of glenoid component positioning.

Material and methods

We studied 10 patients who underwent TSA with use of a PSG to position the glenoid component after preoperative 3D planning. Postoperative glenoid version and tilt were measured and compared to the planned values. We also used new criteria to assess implant rotation and global 3D position, as well as accuracy of the 3D pilot hole for the glenoid guide-pin.

Results

Mean errors in glenoid position were −1.7°±4.4° for version, −0.4°±4.9° for tilt, and 6.0°±13.5° for rotation. Mean difference in global orientation of the glenoid implant versus the planned value was 4.9°±2.5°. Mean 3D discrepancy in glenoid pilot hole position was 2.9±0.5mm; the discrepancy was greater in the mediolateral direction (1.9±0.9mm) than in the supero-inferior (1.1±1.2mm) and antero-posterior (0.8±1.2mm) directions.

Discussion

The poor performance of the PSG in controlling rotation and reaming may explain the difference in global glenoid position compared to the planned value. Improvements in PSG design to incorporate these two parameters deserve consideration.

Level of evidence

II, prospective cohort study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Total shoulder arthroplasty, Patient-specific guide, Accuracy, Glenoid positioning, 3D preoperative planning


Mappa


© 2017  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 104 - N° 1

P. 45-51 - febbraio 2018 Ritorno al numero
Articolo precedente Articolo precedente
  • Recovery after shoulder arthroscopy: Inpatient versus outpatient management
  • M. Boissard, V. Crenn, T. Noailles, S. Campard, F. Lespagnol
| Articolo seguente Articolo seguente
  • Management of the floating shoulder: Does the glenopolar angle influence outcomes? A systematic review
  • V.M. Morey, K.H.Z. Chua, Z.D. Ng, H.M.B. Tan, V.P. Kumar

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.