Accuracy of new transthoracic 3D echocardiographic automated software for left heart chamber quantification in children - 14/08/18

Doi : 10.1016/j.acvdsp.2018.06.004 
Romain Amadieu 1, 2, , Khaled Hadeed 2, Marion Jaffro 3, Clément Karsenty 2, Miarisoa Ratsimandresy 2, Aitor Guitarte Vidaurre 2, Yves Dulac 2, Philippe Acar 2
1 Pediatric Intensive Care Unit, Children's Hospital, CHU Toulouse, France 
2 Department of Pediatric Cardiology, Children's Hospital, CHU Toulouse, France 
3 Department of Radiology, Toulouse University Hospital, CHU Toulouse, France 

Corresponding author.

Riassunto

Introduction

A new three-dimensional echocardiographic (3DE) automated software (HeartModel) is now available to quantify left heart chamber. The aims of this study were to assess the feasibility of this technique in children; and its correlation with manual 3DE and cardiac magnetic resonance (CMR) for measuring left ventricular (LV) and left atrium (LA) volumes and LV ejection fraction (LVEF).

Methods

Ninety-two children (5 to 17 years) were prospectively included in two separate protocols. In protocol, 1, 73 healthy children (8.8±3.0 years) underwent 2D and 3D transthoracic echocardiography (EPIQ 7C, X5-1, Philips Healthcare). LV end-diastolic volume (LVEDV), LV end-systolic volume (LVESV), LVEF and LA volume at ventricular end-systole (LAV) obtained with automated 3DE (Fig. 1) were compared with the manual 3DE measurements. In protocol 2, automated 3DE measurements from 19 children with cardiopathy (12.8±2.9 years) were compared with CMR values. Test-retest, intraobserver and interobserver variability and the mean analysis time per patient were also examined for 3DE measurements.

Results

Automated 3DE was feasible in 77% of datasets and reduced significantly the mean time per patient required for indices analysis compared with manual 3DE (20±2 versus 125±24seconds, P<0.0001), even when contour adjustment was performed (29±10, p<0.0001). In protocol 1, there was excellent correlation for LVEDV, LVESV and LAV between automated and manual 3DE (r=0.89 to 0.99, all P<0.0001) but less for LVEF despite contour adjustment (r=0.56 to 0.57, all P<0.0001). Compared with manual 3DE, automated 3DE without contour edit overestimated LVEDV, LVEF and LAV with small biases and underestimated LVESV with wider bias. With contour adjustment, the biases and limits of agreement (LOA) were reduced (biases: 0.9mL for LVEDV, −1.2mL for LVESV, 2.2% for LVEF; relative biases: 1.3% for LVEDV, −4.5% for LVESV). In protocol 2, there were excellent correlation for LV volumes and moderate correlation for LAV between automated 3DE with contour edit and CMR (r=0.76 to 0.94, all p<0.0003) but the correlation for LVEF remained weak (r=0.46, p=0.05). Compared with CMR, automated 3DE with contour edit slightly underestimated LVEDV and LVESV (relative biases: −7.2 to −7.8%), underestimated LAV with larger bias (relative bias: −31.6%), and had a negligible bias for LVEF (1.0%). However, LOA were clinically acceptable only for LVEDV and LVEF. Test-retest, intraobserver and interobserver variability for automated 3DE measurements were low (<12%).

Conclusions

HeartModel is a promising software for fast assessment of left heart chamber volume and function. Its feasibility in children aged more than 5 years is good, with high reproducibility. The automated 3DE measurements of LV and LA volumes are comparable to manual 3DE, especially when contour adjustment of automated 3DE values is performed. Compared with CMR, LVEDV and LVEF measured by automated 3DE with contour edit seem interesting in clinical practice.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


© 2018  Pubblicato da Elsevier Masson SAS.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 10 - N° 3-4

P. 273 - settembre 2018 Ritorno al numero
Articolo precedente Articolo precedente
  • 15th Annual Meeting of the “Filiale de Cardiologie Pédiatrique et Congénitale”
  • François Godart, Jean-Benoît Thambo
| Articolo seguente Articolo seguente
  • Focal scar and diffuse myocardial fibrosis in patients with history of repaired Tetralogy of Fallot
  • Hubert Cochet, Xavier Iriart, Antoine Allain-Nicolaï, Claudia Camaioni, Soumaya Sridi, Hubert Nivet, Emmanuelle Fournier, Marie-Lou Dinet, Zakaria Jalal, Francois Laurent, Michel Montaudon, Jean-Benoît Thambo

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.