What factors predict ceramic liner malseating after ceramic-on-ceramic total hip arthroplasty? - 21/04/19

Doi : 10.1016/j.otsr.2019.01.009 
Yoshitoshi Higuchi , Taisuke Seki, Yasuhiko Takegami, Yusuke Osawa, Taiki Kusano, Naoki Ishiguro
 Department of Orthopaedic Surgery, Nagoya University Graduate School of Medicine, 65 Tsurumai-cho, Showa-ku, Nagoya, Japan 466-8550 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Background

There is a lack of evidence about the risk factors associated with the malseating of the acetabular liner after ceramic-on-ceramic (CoC) total hip arthroplasty (THA). Therefore, we performed a complementary retrospective case-control study to determine the factors predicting the malseating of the acetabular liner after CoC THA and to evaluate the relationship between malseating and (1) osteoarthritis, (2) particularly in terms of the new radiographic parameter “bone sclerotic length” of the acetabular bone.

Hypothesis

Osteoarthritis, particularly bone sclerotic length, was an independent risk factor for malseating of ceramic liners.

Patients and methods

In total, 219 CoC THAs (174 women and 45 men) were evaluated to determine the risk factors influencing the malseating of the acetabular ceramic liner. An average patient age at the time of surgery was 55.9±9.5 years (range, 23 to 75 years). Data on patient background and preoperative radiographs, such as Tönnis grades; Crowe classification; and indices of acetabular osteoarthritis change; including bone cyst, osteophyte and bone sclerosis, were assessed. The bone sclerotic length in patients with osteoarthritis was measured as the slant distance between the bilateral edges of the sclerosis lesion of the acetabulum on the anteroposterior view.

Results

Preoperative less hip flexion (hazard ratio [HR]: 0.98; 95% CI: 0.97–0.99), osteoarthritis (HR: 3.15; 95% CI: 1.02–9.70) and the bone sclerotic length (HR: 1.83; 95% CI: 1.35–2.48) were independent risk factors determining the malseating of ceramic liners. Receiver operating characteristic curve analysis showed that a bone sclerotic length of 24.6mm was defined as the cut-off point for the malseating of the ceramic liner.

Discussion

Age, preoperative less flextion, osteoarthritis, and the bone sclerotic were independent risk factors determining malseating of ceramic liners. The acetabular shell can also deform upon insertion of the cup with sclerotic bone of the acetabulum and prevent correct seating of liners. Therefore, these factors must be taken into consideration when seating the ceramic liner.

Level of evidence

Case control study III, case control retrospective design.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Ceramic on ceramic bearings, Total hip arthroplasty, Liner malseating, Risk factor, Ceramic liner with a metal-back titanium sleeve


Mappa


© 2019  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 105 - N° 3

P. 453-459 - maggio 2019 Ritorno al numero
Articolo precedente Articolo precedente
  • Does the direct anterior approach allow earlier recovery of walking following total hip arthroplasty? A randomized prospective trial using accelerometry
  • Gautier Bon, Elias Ben Kacem, Pierre Marie Lepretre, Thierry Weissland, Patrice Mertl, Massinissa Dehl, Antoine Gabrion
| Articolo seguente Articolo seguente
  • Dual mobility device reduces the risk of prosthetic hip instability for patients with degenerated spine: A case-control study
  • Louis Dagneaux, Stanislas Marouby, Cedric Maillot, François Canovas, Charles Rivière

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.