Comparison of noise-magnitude and noise-texture across two generations of iterative reconstruction algorithms from three manufacturers - 28/06/19

Doi : 10.1016/j.diii.2019.04.006 
J. Greffier , A. Larbi, J. Frandon, G. Moliner, J.P. Beregi, F. Pereira
 Department of Radiology, CHU Nîmes, University Montpellier, Medical Imaging Group Nîmes, 30029 Nîmes, France 

Corresponding author at: CHU de Nîmes, Medical Imaging Group Nîmes, EA 2415, Bd Prof Robert Debré, 30029 Nîmes Cedex 9, France.CHU de Nîmes, Medical Imaging Group Nîmes, EA 2415Bd Prof Robert DebréNîmes Cedex 930029France

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Purpose

To compare the noise-magnitude and noise-texture across two generations of iterative reconstruction (IR) algorithms proposed by three manufacturers according to the dose level.

Materials and Methods

Five computed tomography (CT) systems equipped with two generations of IR algorithms (hybrid/statistical IR [H/SIR] or full/partial model-based IR [MBIR]) were compared. Acquisitions on Catphan 600 phantom were performed at 120kV and three dose levels (3-, 7- and 12-mGy). Raw data were reconstructed using standard “soft tissue” kernel for filtered back projection and one iterative level of two generations of IR algorithms. Contrast to-noise-ratio (CNR) was computed using three regions of interest: two of them placed in the low-density polyethylene (LDPE) and Teflon® inserts and another placed on the solid water. Noise power spectrum (NPS) was computed to assess the noise-magnitude (NPS peak) and noise-texture (NPS spatial frequency).

Results

CNR increased significantly in MBIR compared to H/SIR algorithms for General-Electric (GE) Healthcare (45%±12 [SD]) and Philips Healthcare systems (62%±11 [SD]) (P<0.001). Regarding Siemens Healthineers systems, CNR of MBIR was significantly lower than that of H/SIR (mean difference: −4%±5 [SD]) (P<0.001) for Teflon® insert but not for LDPE insert (mean difference: −4%±7 [SD]) (P=N.S.). NPS peaks were lower with MBIR than with H/SIR for GE Healthcare (-42%±8 [SD]) and Philips Healthcare (−75%±4 [SD]) systems, whereas it was greater with MBIR than with H/SIR for Siemens Healthineers (13%±11 [SD]) systems. NPS spatial frequencies were higher with MBIR than with H/SIR for Siemens (14%±10 [SD]) but lower for others (−17%±5 [SD] for GE Healthineers and −55%±3 [SD] for Philips Healthcare systems).

Conclusion

This study demonstrates that recent MBIR algorithms, by comparison with the preceding generation, differ according to the main manufacturers with respect to noise-magnitude and noise-texture.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Multidetector computed tomography, Image quality enhancement, Iterative reconstruction, Optimization, Noise power spectrum

Abbreviations : CNR, CT, FBP, GE, H/SIR, IR, MBIR, NPS, ROI


Mappa


© 2019  Société française de radiologie. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 100 - N° 7-8

P. 401-410 - Luglio 2019 Ritorno al numero
Articolo precedente Articolo precedente
  • Interventional radiology techniques for pain reduction and mobility improvement in patients with knee osteoarthritis
  • D. Filippiadis, G. Charalampopoulos, A. Mazioti, E. Alexopoulou, T. Vrachliotis, E. Brountzos, N. Kelekis, A. Kelekis
| Articolo seguente Articolo seguente
  • Predicting TERT promoter mutation using MR images in patients with wild-type IDH1 glioblastoma
  • K. Yamashita, R. Hatae, A. Hiwatashi, O. Togao, K. Kikuchi, D. Momosaka, Y. Yamashita, D. Kuga, N. Hata, K. Yoshimoto, S.O. Suzuki, T. Iwaki, K. Iihara, H. Honda

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2024 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.