Comparative finite element analysis of three implants fixing stable and unstable subtrochanteric femoral fractures: Proximal Femoral Nail Antirotation (PFNA), Proximal Femoral Locking Plate (PFLP), and Reverse Less Invasive Stabilization System (LISS) - 30/01/20

Doi : 10.1016/j.otsr.2019.04.027 
Jie Wang a, b, Jian-Xiong Ma b, Bin Lu b, Hao-Hao Bai b, Ying Wang b, Xin-Long Ma a,
a Department of Orthopaedics, Tianjin Medical University General Hospital, No. 154, Anshan Street, Heping District, 300052, Tianjin China 
b Institute of Orthopaedics, Tianjin Hospital, No. 406, Jiefang Nan Street, Hexi District, 300211 Tianjin, China 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Background

The optimal type of fixation implant for managing subtrochanteric fractures (STFs) is debated, as uncertainty continues to surround the comparative biomechanical performance of the proximal femoral nail antirotation (PFNA), proximal femoral locking plate (PFLP), and reverse less invasive stabilisation system (LISS). No studies have used finite element analysis (FEA) to compare these three devices. The objective of this study was to use FEA to compare the PFNA, PFLP, and LISS used to treat STFs based on the following criteria: (1) stress distribution on the femur and implant, (2) peak stress and stress on the medial side of the femur near the fracture site, and (3) smallest axial displacement of the femoral head.

Hypothesis

Of the three implants, the PFNA has the best biomechanical performance when used for STF fixation.

Methods

FEA was used to assess synthetic bone responses to the three implants used to fix three STF types, namely, Seinsheimer I, III, and IV, characterised by increasing bone loss and/or comminution with subsequent instability. Loading was with 1400N axial compression force.

Results

The LISS and PFLP exhibited similar biomechanical properties in all three fracture types. However, with the Seinsheimer IV fracture, the triangular configuration of the PFLP resulted in stress concentration at the medial and lateral sides of the implant junction. With the Seinsheimer I and III fractures, the PFNA resulted in higher peak stress (183.85MPa and 364.58MPa, respectively) compared to the PFLP (102.90MPa and 177.52MPa) and LISS (116.55MPa and 227.97MPa). With the Seinsheimer IV fracture, peak stress was highest with LISS (2310.40MPa) and was higher with PFLP (2054.90MPa) than with PFNA (1313.30MPa). With the Seinsheimer I and III fractures, the axial femoral head displacement was greater with the PFNA (0.74mm and 1.13mm, respectively) than with the PFLP (0.48mm and 1.02mm) and LISS (0.52mm and 0.92mm). With the Seinsheimer IV fracture, in contrast, the PFNA produced less axial femoral head displacement (4.1mm) compared to the PFLP (12.03mm) and LISS (16.56mm).

Discussion

With unstable (Seinsheimer IV) STFs, fixation stability was better with the PFNA compared to the PFLP and LISS. In contrast, with stable STFs (Seinsheimer I and III), the PFLP and LISS offered greater stability, with similar biomechanical effects. However, with Seinsheimer III fractures, the stress on the implant-femur interface was greater with the LISS.

Level of evidence

IV, basic science study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Subtrochanteric fracture, Finite element analysis, Proximal femoral nail antirotation, Less invasive stabilisation system, Proximal femoral locking plate


Mappa


© 2019  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 106 - N° 1

P. 95-101 - febbraio 2020 Ritorno al numero
Articolo precedente Articolo precedente
  • Application of the Radiographic Union Scale for Tibial fractures (RUST): Assessment of healing rate and time of tibial fractures managed with intramedullary nailing
  • Jun Min Leow, Nicholas D. Clement, A. Hamish W.R. Simpson
| Articolo seguente Articolo seguente
  • Annual case volume is a risk factor for 30-day unplanned readmission after open reduction and internal fixation of acetabular fractures
  • Robert Kent Merrill, Blake Ryan Turvey, Germanuel Lavar Landfair, Emmanuel Michael Illical

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.