Abbonarsi

Colonic endoscopic submucosal dissection using a novel robotic system (with video) - 17/04/21

Doi : 10.1016/j.gie.2020.09.042 
Philip Wai Yan Chiu, MD (CUHK), FRCSEd, FASGE 1, , Khek Yu Ho, MD, FRACP, FRCP 2, Soo Jay Phee, PhD 3
1 Department of Surgery, Institute of Digestive Disease, Faculty of Medicine, The Chinese University of Hong Kong, Hong Kong, China 
2 Division of Gastroenterology, Department of Medicine, National University of Singapore, Singapore 
3 School of Mechanical & Aerospace Engineering, College of Engineering, Nanyang Technological University, Singapore 

Reprint requests: Philip Chiu, Division of Upper GI and Metabolic Surgery, Department of Surgery, Faculty of Medicine, The Chinese University of Hong Kong, 30-32 Ngan Shing St Shatin, Hong Kong 00000, Hong Kong.Division of Upper GI and Metabolic SurgeryDepartment of SurgeryFaculty of MedicineThe Chinese University of Hong Kong30-32 Ngan Shing St ShatinHong Kong00000Hong Kong

Abstract

Background and Aims

One of the difficulties in performing endoscopic submucosal dissection (ESD) is the lack of retraction during submucosal dissection. The development of the EndoMaster EASE System (EndoMaster Pte Ltd, Singapore) aims to enhance the safety and efficacy of ESD through 2 flexible robotic arms for tissue retraction and dissection. This is a preclinical animal study to evaluate the performance of colorectal ESD using the latest version of the EndoMaster EASE System.

Methods

The latest version of the EndoMaster EASE System consists of an independently designed, flexible platform with a built-in endoscopic imaging system and 3 working channels, 2 for the passage of robotic arms and 1 for accessories. In this animal study, the outcome measures were operating time (from starting incision to finishing dissection), completeness of resection, procedure-related adverse events, and limitations of arm manipulation in a narrow working space as assessed by counting the frequency of blind cutting.

Results

Five ESD procedures were performed in a 66.7-kg porcine model with the animal under general anesthesia. The mean operative time was 73.8 minutes, and the mean size of the specimen resected was 1340 mm2. There was no perforation, although profuse bleeding was encountered during 1 robotic ESD procedure.

Conclusions

The current preclinical study confirmed the feasibility of performing colorectal ESD using the latest version of the EndoMaster EASE System. The system was also tested for the ability to manage adverse events including bleeding and perforation. This study provided important preclinical experience for clinical trial.

Il testo completo di questo articolo è disponibile in PDF.

Abbreviation : ESD


Mappa


 DISCLOSURE: The following authors disclosed financial relationships: P. W. Y. Chiu: Advisory board member for EndoMaster Pte Ltd. K. Y. Ho, S. J. Phee: Co-founders of EndoMaster Pte Ltd. Research support for this study was provided by EndoMaster Pte Ltd.
 If you would like to chat with an author of this article, you may contact Dr Chiu at philipchiu@surgery.cuhk.edu.hk.


© 2021  American Society for Gastrointestinal Endoscopy. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 93 - N° 5

P. 1172-1177 - maggio 2021 Ritorno al numero
Articolo precedente Articolo precedente
  • Continuing Medical Education Exam: May 2021
  • Karthik Ravi, William Ross, Ara Sahakian, Brian Weston, Prasad G. Iyer, Amit Rastogi, Michael B. Wallace
| Articolo seguente Articolo seguente
  • Expanding a new frontier: endoscopic submucosal dissection by use of a robotic endoluminal system
  • Sehrish Jamot, Richard Rothstein

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.