Needle artifact characteristics and insertion accuracy using a 1.2 T open MRI scanner: A phantom study - 26/05/21

Doi : 10.1016/j.diii.2020.12.007 
Soichiro Okamoto a, Yusuke Matsui a, , Takao Hiraki a, Toshihiro Iguchi a, Toshiyuki Komaki a, Takatsugu Yamauchi b, Mayu Uka a, Koji Tomita a, Jun Sakurai c, Hideo Gobara d, Susumu Kanazawa a
a Department of Radiology, Okayama University Medical School, 2-5-1 Shikatacho, Kitaku, Okayama 700-8558, Japan 
b Central Division of Radiology, Okayama University Hospital, 2-5-1 Shikatacho, Kitaku, Okayama 700-8558, Japan 
c Center for Innovative Clinical Medicine, Okayama University Hospital, 2-5-1 Shikatacho, Kitaku, Okayama 700-8558, Japan 
d Division of Medical Informatics, Okayama University Hospital, 2-5-1 Shikatacho, Kitaku, 700-8558, Okayama, Japan 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Highlights

Clinical use of a 1.2 Tesla open system for MRI-guided intervention is awaited.
Needle artifact characteristics using a 1.2 Tesla open MRI system are clarified using a phantom study.
Accuracy of needle insertion using a 1.2 Tesla open MRI system is acceptable for clinical use.

Il testo completo di questo articolo è disponibile in PDF.

Abstract

Purpose

To evaluate the characteristics of needle artifacts and the accuracy of needle insertion using a 1.2 Tesla open magnetic resonance imaging (MRI) system in a phantom.

Materials and methods

First, the apparent width of the needle on the MRI and the needle tip position error of 16- and 18-gauge MRI-compatible introducer needles and a 17-gauge cryoneedle were examined with different needle angles (0°, 30°, 45°, 60°, and 90°) to the main magnetic field (B0), sequence types (balanced steady-state acquisition with rewound gradient echo [BASG] and T2-weighted fast spin echo [FSE] sequence), and frequency encoding directions. Second, the accuracy of needle insertion was evaluated after 10 MRI fluoroscopy-guided insertions in a phantom.

Results

The apparent needle widths was larger when the angle of the needle axis relative to B0 was larger. The needles appeared larger on BASG than on T2-weighted FSE images, with the largest apparent widths of 16-, 17-, and 18-gauge needles of 14.3, 11.6, and 11.0mm, respectively. The apparent needle tip position was always more distal than the actual position on BASG images, with the largest longitudinal error of 4.0mm. Meanwhile, the 16- and 18-gauge needle tips appeared more proximal on T2-weighted FSE images with right-to-left frequency encoding direction. The mean accuracy of MRI fluoroscopy-guided needle insertion was 3.1mm.

Conclusion

These experiments clarify the characteristics of needle artifacts in a 1.2 Tesla open MRI. With this system, the MRI fluoroscopy-guided needle insertion demonstrated an acceptable accuracy for clinical use.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Magnetic resonance imaging, Needles, Artifacts, Phantoms, Imaging


Mappa


© 2021  Société française de radiologie. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 102 - N° 6

P. 363-370 - giugno 2021 Ritorno al numero
Articolo precedente Articolo precedente
  • Percutaneous thermal ablation of sacral metastases: Assessment of pain relief and local tumor control
  • Roberto L. Cazzato, Pierre De Marini, Ian Leonard-Lorant, Danoob Dalili, Guillaume Koch, Pierre A. Autrusseau, Theo Mayer, Julia Weiss, Pierre Auloge, Julien Garnon, Afshin Gangi
| Articolo seguente Articolo seguente
  • CT features of lung metastases from pancreatic adenocarcinoma: Correlation with histopathologic findings
  • Mathilde Aissaoui, Audrey Lupo, Romain Coriat, Benoit Terris, Souhail Bennani, Guillaume Chassagnon, Marie-Pierre Revel

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.