New image quality and dose reduction technique for pediatric digital radiography - 26/06/21

Doi : 10.1016/j.diii.2021.01.009 
Joelle A. Feghali a, ⁎ , Greg Chambers a, Julie Delépierre a, Sophie Chapeliere a, Inès Mannes a, b, Catherine Adamsbaum a, b
a AP–HP, Bicêtre Hospital, Pediatric Radiology Department, 94270 Le Kremlin-Bicêtre, France 
b Medical School, Paris-Saclay University, 94270 Le Kremlin-Bicêtre, France 

⁎Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Highlights

•
A novel approach to optimise pediatric digital radiography (DR) was introduced using a simulation tool and a specific image quality grid.
•
S-Vue™ image denoising was applied to thorax, pelvis, abdomen and spine DR images and its effect was assessed through simulation tool and image quality grid.
•
S-Vue™ can yield digital radiography images of pediatric patients with high diagnostic quality at half the standard radiation dose.

Il testo completo di questo articolo è disponibile in PDF.

Abstract

Purpose

The purpose of this study was to test a new post-processing and denoising engine for patient dose reduction while maintaining diagnostic image quality (IQ) in pediatric digital radiography (DR).

Materials and methods

Pediatric DR images of the thorax, pelvis, abdomen and spine obtained in 174 patients (102 males, 72 females; mean age, 2±1.8 [SD] years; age range: 6 months–9 years) were retrieved. Artificial noise was added to the images to simulate acquisitions at 50%, 32% and 12.5% of the routine dose levels. A total of 696 images corresponding to four dose levels were post-processed using S-Vue™ and further blindly scored by three pediatric radiologists using a scoring grid of 4–6 criteria specifically defined per anatomical area. The mean score was assessed for each area and weight class (5–15 and 15–30kg) and compared across the simulated low dose images. Paired Wilcoxon test was used with a threshold difference of 0.5 (half a criterion) between scores to highlight a significant reduction in image quality. Inter-rater reliability was assessed using intraclass correlation coefficient (ICC).

Results

Only the 50% reduced dose images showed non-inferiority when compared to routine images for all of areas and weight classes (P<0.01). Very good inter-rater reliability of the overall scores was observed for the pelvis in the 5–15kg weight class (ICC=0.85) for images at full dose, 50% and 32% reduced dose. For the remaining areas (thorax, abdomen and spine) and weight classes, inter-rater reliability was moderate (ICC: 0.3–0.6).

Conclusion

S-Vue™ post-processing software allows a two-fold radiation dose reduction while maintaining satisfactory IQ in pediatric DR.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Digital radiography, Pediatrics, Dose reduction, Image processing, Image noise

Abbreviations : ALADA, ALARA, CI, CR, DR, DRL, ICC, IQ, SD


Mappa


© 2021  Société française de radiologie. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 102 - N° 7-8

P. 463-470 - luglio 2021 Ritorno al numero
Articolo precedente Articolo precedente
  • MRI-based traditional radiomics and computer-vision nomogram for predicting lymphovascular space invasion in endometrial carcinoma
  • Ling Long, Jianqing Sun, Liling Jiang, Yixin Hu, Lan Li, Yong Tan, Meimei Cao, Xiaosong Lan, Jiuquan Zhang
| Articolo seguente Articolo seguente
  • Utility of enteric contrast material in abdominal penetrating trauma: A narrative review
  • Victoria F. Tan, Vincent M. Mellnick, Michael N. Patlas

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.