Reduced nontarget embolization and increased targeted delivery with a reflux-control microcatheter in a swine model - 23/09/21

Doi : 10.1016/j.diii.2021.05.002 
Silvia Rizzitelli a, 1, , Nir Holtzman b, 1, Geert Maleux c, Thierry De Baere d, Fei Sun e, Pierre-Olivier Comby f, Michael Tal g, Gwenaelle Bazin a, Francois Montestruc h, Thomas Viel i, Philippe Robert a, Osnat Harbater b, Eran Miller b, Claire Corot a
a R&I, Guerbet, 95943 Roissy CdG Cedex, France 
b Guerbet Company, Accurate Medical Therapeutics LTD, 7670203 Rehovot, Israel 
c Radiology, University Hospitals Leuven, 3000 Leuven, Belgium 
d Interventional Radiology, Institut Gustave Roussy, 94800 Villejuif, France 
e Endoluminal Therapy and Diagnosis Unit, Jesús Usón Minimally Invasive Surgery Centre, 10071 Cáceres, Spain 
f Department of Neuroradiology, François-Mitterrand University Hospital, 21079 Dijon, France 
g Interventional Radiology, Hadassah Hospital, 9112001 Jerusalem, Israel 
h MSC, eXYSTAT, 92240 Malakoff, France 
i Life Imaging Facility (PIV), Université de Paris, PARCC-INSERM U970, 75015 Paris, France 

Corresponding author at: R&I, Guerbet, 95943 Roissy CdG Cedex, France.R&I, GuerbetRoissy CdG Cedex95943France

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Highlights

A reflux control microcatheter was compared to a standard end hole microcatheter in terms of non-target embolization in a swine model.
A reflux control microcatheter delivers 98% of microspheres in the target site of injection compared to 89% with a standard end-hole microcatheter.
Percentage of non-target embolization calculated by micro computed tomography analysis is greater with a standard end hole microcatheter that with a reflux control microcatheter.
Qualitative observation of micro computed tomography images demonstrates an overall superiority of the reflux-control microcatheter compared to a standard end-hole microcatheter for renal embolization with radiopaque microspheres in a swine model.

Il testo completo di questo articolo è disponibile in PDF.

Abstract

Purpose

To evaluate the potential differences in non-target embolization and vessel microsphere filling of a reflux-control microcatheter (RCM) compared to a standard end-hole microcatheter (SEHM) in a swine model.

Materials and methods

Radiopaque microspheres were injected with both RCM and SEHM (2.4-Fr and 2.7-Fr) in the kidneys of a preclinical swine model. Transarterial renal embolization procedures with RCM or SEHM were performed in both kidneys of 14 pigs. Renal arteries were selectively embolized with an automated injection protocol of radio-opaque microspheres. Ex-vivo X-ray microtomography images of the kidneys were utilized to evaluate the embolization by quantification of the deposition of injected microspheres in the target vs. the non-target area of injection. X-ray microtomography images were blindly analyzed by five interventional radiologists. The degree of vessel filling and the non-target embolization were quantified using a scale from 1 to 5 for each parameter. An analysis of variance was used to compare the paired scores.

Results

Total volumes of radio-opaque microspheres injected were similar for RCM (11.5±3.6 [SD] mL; range: 6–17mL) and SEHM (10.6±5.2 [SD] mL; range: 4–19mL) (P=0.38). The voxels enhanced ratio in the target (T) vs. non-target (NT) areas was greater with RCM (T=98.3% vs. NT=1.7%) than with SEHM (T=89% vs. NT=11%) but the difference was not significant (P=0.30). The total score blindly given by the five interventional radiologists was significantly different between RCM (12.3±2.1 [SD]; range: 6–15) and the standard catheter (11.3±2.5 [SD]; range: 4–15) (P=0.0073), with a significant decrease of non-target embolization for RCM (3.8±1.3 [SD]; range: 3.5–4.2) compared to SEHM (3.2±1.5 [SD]; range: 2.9–3.5) (P=0.014).

Conclusion

In an animal model, RCM microcatheters reduce the risk of non-target embolization from 11% to 1.7%, increasing the delivery of microspheres of 98% to the target vessels, compared to SEHM microcatheters.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Embolization, Therapeutic, Interventional radiology, Non-target embolization, X-ray microtomography, Preclinical evaluation

Abbreviations : CI, LS means, μCT, NTE, RCM, SEHM, SD, TACE


Mappa


© 2021  The Author(s). Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 102 - N° 10

P. 641-648 - ottobre 2021 Ritorno al numero
Articolo precedente Articolo precedente
  • Diffusion-weighted MRI and PET/CT reproducibility in epithelial ovarian cancers during neoadjuvant chemotherapy
  • Amandine Crombé, Lisa Gauquelin, Stéphanie Nougaret, Marine Chicart, Marina Pulido, Anne Floquet, Frédéric Guyon, Sabrina Croce, Michèle Kind, Anne-Laure Cazeau
| Articolo seguente Articolo seguente
  • Imaging of Oxford/AstraZeneca® COVID-19 vaccine-induced immune thrombotic thrombocytopenia
  • Matthieu Garnier, Adelya Curado, Paul Billoir, Virginie Barbay, Matthieu Demeyere, Jean-Nicolas Dacher

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.