Predicting difficult intubation: the hyomental distance ultrasound evaluation is superior to the thyromental distance - 24/11/22

Doi : 10.1016/j.accpm.2022.101144 
Bin Wang a, 1, Mingfang Wang a, 1, Fangfang Yang a, Changjian Zheng a, Tao Yu b, Jianling Xu a, Yongquan Chen a, Weidong Yao a, b,
a Department of Anaesthesiology, Yijishan Hospital of Wannan Medical College, Wuhu, Anhui, China 
b Anhui Province Clinical Research Center for Critical Care Medicine (Respiratory Disease), Yijishan Hospital of Wannan Medical College, Wuhu, Anhui, China 

Corresponding author.

Abstract

Background

Hyomental distance (HMD), an anatomical feature of the upper airway, can be measured precisely by ultrasonography. But the sensitivity and specificity of HMD compared to thyromental distance (TMD) to predict difficult airways is still unknown.

Methods

A case-cohort study was conducted. The written informed consent was obtained. Elective surgery adult patients undergoing general anaesthesia and tracheal intubation were recruited. The other inclusion criteria were: no maxillofacial deformity, trauma, airway stenosis, known difficult airway. The exclusion criteria were: tracheal intubations or operations were canceled, or patients’ data were missing. HMD ultrasound measurements of patients in a sniffing position and other usual airway evaluations were performed before general anaesthesia induction. The primary outcome was the intubation difficulty level. Predictive values were calculated.

Results

We successfully enrolled 2357 patients (62 difficult intubation patients) in the cohort study for analysis. The area under the receiver operating characteristic curve (AUC) of the HMD and TMD for predicting difficult intubation was 0.86 (95% CI, 0.84−0.87) and 0.77 (95% CI, 0.75−0.78) respectively (comparison: P < 0.001). With an optimal cut-off value of HMD ≤ 4.9 cm, we observed a sensitivity and specificity of 0.90 (95% CI, 0.80−0.96) and 0.73 (95% CI, 0.71−0.75). Meanwhile, with TMD ≤ 7.0 cm, the sensitivity and specificity were 0.77 (95% CI, 0.65−0.87) and 0.65 (95% CI, 0.63−0.67) respectively.

Conclusion

In comparison to TMD, HMD measured by ultrasound was more sensitive in predicting difficult intubation.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Difficult intubation, Hyomental distance, Ultrasound, Thyromental distance, Airway management


Mappa


© 2022  Société française d'anesthésie et de réanimation (Sfar). Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 41 - N° 6

Articolo 101144- dicembre 2022 Ritorno al numero
Articolo precedente Articolo precedente
  • Automated systems to minimise asynchronies and personalise mechanical ventilation: A light at the end of the tunnel!
  • Paolo Pelosi, Lluis Blanch, Matthieu Jabaudon, Jean-Michel Constantin
| Articolo seguente Articolo seguente
  • Comparison of two electromyography-based neuromuscular monitors, AF-201P and TetraGraph, in rocuronium-induced neuromuscular block: A prospective comparative study
  • Hanae Sato, Hajime Iwasaki, Akira Doshu-Kajiura, Seidai Katagiri, Shunichi Takagi, Sarah Kyuragi Luthe, Takahiro Suzuki

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.