Accuracy and variability of cardiologist interpretation of single lead electrocardiograms for atrial fibrillation: The VITAL-AF trial - 11/10/23

Doi : 10.1016/j.ahj.2023.07.003 
Daniel C. Pipilas, MD a, b, Shaan Khurshid, MD, MPH a, b, c, Steven J. Atlas, MD, MPH b, d, Jeffrey M. Ashburner, PhD, MPH b, d, Ana T. Lipsanopoulos, BS a, Leila H. Borowsky, MPH d, Wyliena Guan, MS e, Patrick T. Ellinor, MD, PhD a, b, c, David D. McManus, MD, MSc f, Daniel E. Singer, MD b, d, Yuchiao Chang, PhD b, d, Steven A. Lubitz, MD, MPH b, c,
a Division of Cardiology, Massachusetts General Hospital, Boston, MA 
b Harvard Medical School, Boston, MA 
c Demoulas Center for Cardiac Arrhythmias and Cardiovascular Research Center, Massachusetts General Hospital, Boston, MA 
d Division of General Internal Medicine, Massachusetts General Hospital, Boston, MA 
e University of North Carolina, Chapel Hill, NC 
f Department of Medicine, University of Massachusetts Medical School, Worcester, MA 

Reprint requests: Steven A. Lubitz, MD, MPH, Demoulas Center for Cardiac Arrhythmias and Cardiovascular Research Center, Massachusetts General Hospital, 55 Fruit Street, GRB 109, Boston, MA, 02114.Demoulas Center for Cardiac Arrhythmias and Cardiovascular Research CenterMassachusetts General Hospital, 55 Fruit Street, GRB 109BostonMA02114

Riassunto

Background

Screening for atrial fibrillation (AF) using consumer-based devices capable of producing a single lead electrocardiogram (1L ECG) is increasing. There are limited data on the accuracy of physician interpretation of these tracings. The goal of this study is to assess the sensitivity, specificity, confidence, and variability of cardiologist interpretation of point-of-care 1L ECGs.

Methods

Fifteen cardiologists reviewed point-of-care handheld 1L ECGs collected from patients aged 65 years or older enrolled in the VITAL-AF clinical trial [NCT035115057] who underwent cardiac rhythm assessments with a 1L ECG using an AliveCor KardiaMobile device. Random sampling of 1L ECGs for cardiologist review was stratified by the AliveCor algorithm interpretation. A 12L ECG performed on the same day for clinical purposes was used as the gold standard. Cardiologists each reviewed a common sample of 200 1L ECG tracings and completed a survey associated with each tracing. Cardiologists were blinded to both the AliveCor algorithm and same day 12L ECG interpretation. For each tracing, study cardiologists were asked to assess the rhythm (sinus rhythm, AF, unclassifiable), report their assessment of the quality of the tracing, and rate their confidence in rhythm interpretation. The outcomes included the sensitivity, specificity, variability, and confidence in physician interpretation. Variables associated with each measure were identified using multivariable regression.

Results

The average sensitivity for AF was 77.4% (range 50%-90.6%, standard deviation [SD]=11.4%) and the average specificity was 73.0% (range 41.3%-94.6%, SD = 15.4%). The mean variability was 30.8% (range 0%-76.2%, SD = 23.2%). The average reviewer confidence of 1L ECG rhythm assessment was 3.6 out of 5 (range 2.5-4.2, SD = 0.6). Patient and tracing factors associated with sensitivity, specificity, variability, and confidence were identified and included age, body mass index, and presence of artifact.

Conclusion

Cardiologist interpretation of point-of-care handheld 1L ECGs has modest diagnostic sensitivity and specificity with substantial variability for AF classification despite high confidence. Variability in cardiologist interpretation of 1L ECGs highlights the importance of confirmatory testing for diagnosing AF.

Il testo completo di questo articolo è disponibile in PDF.

Graphical Abstract




Image, graphical abstract

Il testo completo di questo articolo è disponibile in PDF.

Mappa


© 2023  Elsevier Inc. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 265

P. 92-103 - novembre 2023 Ritorno al numero
Articolo precedente Articolo precedente
  • Prognostic impact of precipitated cardiac decompensation in symptomatic heart failure with reduced ejection fraction and severe secondary mitral regurgitation
  • Nathan Mewton, Erwan Donal, François Picard, François Derimay, Daniel Grinberg, Delphine Maucort Boulch, Thomas Bochaton, Nicolas Piriou, Amélie De Lorgeril, Geraldine Samson, Frédéric Rouleau, Benjamin Riche, Jean Noël Trochu
| Articolo seguente Articolo seguente
  • Prognostic value of coronary artery calcium score for the prediction of atherosclerotic cardiovascular disease in participants with suspected nonalcoholic hepatic steatosis: Results from the multi-ethnic study of atherosclerosis
  • Keishi Ichikawa, Spencer Hansen, Venkat S. Manubolu, Leili Pourafkari, Hooman Fazlalizadeh, Jairo Aldana-Bitar, Lisa B. VanWagner, Srikanth Krishnan, Matthew J. Budoff

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.