Using a patient-specific cutting guide enables identical knee osteotomies: An evaluation of accuracy on sawbones - 27/11/24

Doi : 10.1016/j.otsr.2024.103813 
Sébastien Avellan a, b, c, Ahmed Mabrouk b, d, Virginie Taillebot a, b, , Martine Pithioux a, b, Matthieu Ollivier a, b
a Aix Marseille Univ, CNRS, ISM, 13009 Marseille, France 
b Aix Marseille Univ, APHM, CNRS, ISM, Sainte-Marguerite Hospital, Institute for Locomotion, Department of Orthopaedics and Traumatology, 13009 Marseille, France 
c BIOBank®, Tissue Bank, Lieusaint, France 
d Leeds Teaching Hospitals, Leeds, United Kingdom 

Corresponding author. Aix-Marseille University, 163, avenue de Luminy, 13009 Marseille, France. Aix-Marseille University 163, avenue de Luminy Marseille 13009 France

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Purpose

It was hypothesized that using a Patient-Specific Cutting Guide (PSCG) would allow the creation of sawbones model osteotomies, identical in the 3 planes and the hinge parameters, that can be used for biomechanical studies. The aim of the study was to evaluate the accuracy of the PSCG system and to introduce and assess the new hinge parameter; the hinge area.

Methods

Six identical sawbones tibia models were identically set up for identical osteotomy cuts by the same surgeon in the same session and with identical instruments. A medical scanner was used to evaluate the 3D configuration of all the specimens. The analyzed parameters included the cutting angles in both the coronal and sagittal planes (degrees) and the hinge and the slicing areas (cm 2 ), and the hinge thickness (mm). The values were statistically evaluated for average, standard deviation, 95% confidence index, and delta to the expected values were calculated.

Results

The mean values for the coronal and sagittal angles were 110.5̊ ± 1̊ and 89.8̊ ± 0.8̊, respectively. The 95% confidence index level ranged between 0.1̊, and 0.8̊ in both the coronal & the sagittal planes. The mean values for the hinge thickness, the hinge area, and the slicing area were 12.7 ± 1.5 mm, 4.2 ± 0.9 cm 2 , and 18.3 ± 1.2 cm 2 , respectively.

Conclusion

In the presented study, it can be demonstrated that mechanically identical osteotomy specimens, with regard to the cutting planes and hinge parameters, can be reliably created using the PSCG. The identical specimens can be used for biomechanical research purposes to further expand our knowledge of the factors affecting osteotomy outcomes.

Level of evidence

IV.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Patient-specific instrumentation, Cutting guide, Cut accuracy-exactness, Precision, Trueness, High tibial osteotomy, Bone composite, Sawbones, Hinge


Mappa


© 2024  The Authors. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 110 - N° 8

Articolo 103813- dicembre 2024 Ritorno al numero
Articolo precedente Articolo precedente
  • In vitro comparative study of deformation of 3D-printed models using different polylactic acids treated by steam sterilization
  • Antoine Moreau, Louis Rony, Antoine Robelet, Hélène Laubacher, Anne-Valérie Lebelle-Dehaut
| Articolo seguente Articolo seguente
  • Computer aided multiplanar osteotomy using patient specific instrumentation to treat cubitus varus in children
  • Estelle Alonso, Christian Victoria, Nizar Touati, Raphael Vialle, Franck Fitoussi, Manon Bachy

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.