Total knee arthroplasty three-dimensional kinematic estimation prevision. From a two-dimensional fluoroscopy acquired dynamic model - 09/04/11

Doi : 10.1016/j.otsr.2011.01.003 
B.P. Lebel a, , V. Pineau a, S.L. Gouzy a, L. Geais b, J.-J.M. Parienti c, J.-J.P. Dutheil c, C.H. Vielpeau a
a Orthopedic Department, Caen University Hospital, avenue de la Côte-de-Nacre, 14033 Caen cedex, France 
b Reasearch and Development, Stryker SA, Europe 
c Biostatistic Department, Caen University Hospital, Caen, France 

Corresponding author. Tel.: +33 2 31 06 48 09; fax: +33 23 10 64 85 09.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Summary

Introduction

To determine six-degree of freedom of total knee arthroplasty kinematics (TKA), optimized matching algorithms for single fluoroscopic image system may be used. Theoretical accuracy of these systems was reported. Nevertheless, all reports were done under idealized laboratory experimental conditions. The aim of this study was to evaluate the “true” accuracy of a flat panel single plane video-fluoroscopy system based on computed-assisted design (CAD) model matching and compare it to TKA kinematics obtained from optoelectronic measurements as gold standard.

Hypothesis

The estimation of the error produced by 2D/3D fluoroscopic registration in daily practice is misjudged in most available laboratory reports.

Material and methods

The experimental set-up used a TKA implanted into femoral and tibial cadaver bones. Thirty flexions were simultaneously registered using single plane fluoroscopy and an active optical tracking system. Kinematics registered were compared using the root mean square error (RMS), the concordance correlation coefficient and Bland & Altman plot analysis.

Results

The mean range of motion of flexion during the experiment was 106°. The respective RMS for flexion, varus-valgus and internal-external rotation were 0.68, 0.67 and 1.02°. The respective RMS for antero-posterior, medio-lateral and proximo-distal displacement were 1.3, 2.4 and 1.06mm. Extreme values of the measured error concerning medio-lateral displacement were −5.4 and 22,1mm.

Discussions

Analysis found some outliners in all degree of freedom with a systematic error and larger standard deviation than already published data. One should make sure that during the experiment the motion of interest is in the in-plane direction. Moreover, this study brings out the true threshold detection of this type of analysis.

Level of evidence

Level 3. Prospective diagnostic study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Kinematics, Total knee arthroplasty, Fluoroscopy, Image processing, Optoelectronic, Accuracy


Mappa


© 2011  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 97 - N° 2

P. 111-120 - aprile 2011 Ritorno al numero
Articolo precedente Articolo precedente
  • Contents
| Articolo seguente Articolo seguente
  • Stem subsidence after total hip revision: 183 cases at 5.9 years follow-up
  • J. Girard, O. Roche, G. Wavreille, F. Canovas, P. Le Béguec

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.