Abbonarsi

Component impingement in total hip arthroplasty: Frequency and risk factors. A continuous retrieval analysis series of 416 cup - 09/04/11

Doi : 10.1016/j.otsr.2010.12.004 
E. Marchetti a, b, N. Krantz a, b, C. Berton a, b, D. Bocquet a, b, N. Fouilleron a, b, H. Migaud a, b, , J. Girard a, b
a Lille–Nord-de-France University, 59000 Lille, France 
b Orthopaedic department, Roger-Salengro Hospital, Lille University Hospital, Place de Verdun, 59037 Lille, France 

Corresponding author. Tel.: +33 0320 446 828; fax: +33 0320 446 607.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Summary

Introduction

Impingement is a factor of failure in total hip replacement (THR), causing instability and early wear. Its true frequency is not known; cup-retrieval series reported rates varying from 27 to 84%.

Hypothesis

The hypothesis was that a large continuous series of THR cup removals would help determine the frequency of component impingement.

Objectives

The hypothesis was tested on a continuous retrospective series of cups removed in a single center, with a secondary objective of identifying risk factors.

Materiel and methods

Macroscopic examination looked for component impingement signs in 416 cups retrieved by a single operator between 1989 and 2004. Risk factors were investigated by uni- and multivariate analyses in the 311 cases for which there were complete demographic data. In these 311 cases, removal was for aseptic loosening (131 cases), infection (43 cases), instability (56 cases), osteolysis (28 cases) or unexplained pain (48 cases); impingement was explicitly implicated in only five cases (1.6%), always with hard-on-hard bearing components.

Results

Impingement was found in 214 of the 416 cups (51.4%) and was severe (notch>1mm) in 130 (31.3%). In the subpopulation of 311 cups, impingement was found in 184 cases (59.2%) and was severe in 109 (35%). Neither duration of implant use nor cup diameter or frontal orientation emerged as risk factors. On univariate analysis, impingement was more frequently associated with revision for instability, young patient age at THR, global hip range of motion >200° or use of an extended femoral head flange (or of an elevated antidislocation rim liner), and was more severe in case of head/neck ratio<2. On multivariate analysis, only use of an extended head flange (RR 3.2) and revision for instability (RR 4.2) remained as independent risk factors for impingement.

Discussion

Component impingement is frequently observed in cups after removal, but is rarely found as a direct indication for revision, except in case of hard-on-hard friction couples (polyethylene being the most impingement-tolerant material). Systematic use of extended head flanges and elevated antidislocation rims is not to be recommended, especially in case of excessive ROM. A good head/neck ratio should be sought, notably by increasing the head diameter in less impingement-tolerant hard-on-hard friction couples. Although not identified as a risk factor in the present study, implant orientation should be checked; computer-assisted surgery can be useful in this regard, for adaptation to the patient’s individual range-of-motion cone.

Level of evidence

Level IV, retrospective observational study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Impingement, Instability, Wear, Hip prosthesis, Dislocation, Retrieval analysis


Mappa


© 2011  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 97 - N° 2

P. 127-133 - aprile 2011 Ritorno al numero
Articolo precedente Articolo precedente
  • Stem subsidence after total hip revision: 183 cases at 5.9 years follow-up
  • J. Girard, O. Roche, G. Wavreille, F. Canovas, P. Le Béguec
| Articolo seguente Articolo seguente
  • Infected total hip arthroplasty treated by an irrigation-debridement/component retention protocol. A prospective study in a 12-case series with minimum 2 years’ follow-up
  • S. Klouche, L. Lhotellier, P. Mamoudy

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.