Abbonarsi

Early Experience with a Simplified Technique for Transcatheter Closure of the Patent Foramen Ovale - 12/08/11

Doi : 10.1016/j.hlc.2009.08.005 
Shelby Kutty, MD a, David G. Nykanen, MD b, Evan M. Zahn, MD b,
a University of Nebraska/Creighton University Joint Division of Pediatric Cardiology, Children’s Hospital and Medical center, Omaha, NE, United States 
b Division of Cardiology, The Congenital Heart Institute at Miami Children’s Hospital and Arnold Palmer Hospital, Miami and Orlando, FL, United States 

Corresponding author at: Miami Children’s Hospital, 3100 SW 62nd Avenue, Miami, FL 33155, United States.

Riassunto

Background

Although generally safe, transcatheter patent foramen ovale (PFO) closure typically involves catheter exchanges and manipulations within the left atrium (LA) that may result in complications. We describe the development of a simplified technique for PFO closure, further to modification of our earlier approaches.

Methods

Early in our experience, LA angiography and/or balloon sizing of the PFO was routine prior to device closure. Subsequently, the simplified technique described herein became our standard approach. We reviewed immediate and medium term results in 27 consecutive patients who underwent closure in a 24-month time frame.

Results

All procedures used general anaesthesia, TEE guidance and CardioSEAL septal occluder. Device delivery was successful in all patients. The median fluoroscopy time was significantly shorter with the simplified technique (median difference 6.7min, p<0.05) with a 98.3% median confidence interval (4.5, 19.9min). There were four complications, all occurring in the early group of patients. There have been no recurrent neurological events or device related adverse events on medium term follow-up.

Conclusions

Successful transcatheter PFO closure is feasible without LA angiography or catheter-based balloon sizing, utilising echocardiographic guidance for device selection and placement. Simplification of PFO closure technique could potentially decrease complications and shorten fluoroscopy times.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Patent foramen ovale, Paediatric cardiology, Adult congenital heart disease, Interventional catheterisation


Mappa


© 2009  Australasian Society of Cardiac and Thoracic Surgeons and the Cardiac Society of Australia and New Zealand. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 18 - N° 6

P. 384-387 - dicembre 2009 Ritorno al numero
Articolo precedente Articolo precedente
  • Post-cardiac injury syndrome following transvenous pacemaker insertion: A case report and review of the literature
  • Cihan Cevik, Troy Wilborn, Rita Corona, Elizabeth Schanzmeyer, Kenneth Nugent
| Articolo seguente Articolo seguente
  • The COACH Program Produces Sustained Improvements in Cardiovascular Risk Factors and Adherence to Recommended Medications—Two Years Follow-up
  • Michael Jelinek, Margarite J. Vale, Danny Liew, Leeanne Grigg, Anthony Dart, David L. Hare, James D. Best

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.