Effect of defibrillation testing on management during implantable cardioverter-defibrillator implantation - 18/08/11

Doi : 10.1016/j.ahj.2006.06.025 
Peter Leong-Sit, MD FRCPC, Lorne J. Gula, MD MSc FRCPC , Pantelis Diamantouros, MD FRCPC, Andrew D. Krahn, MD FRCPC FACC, Allan C. Skanes, MD FRCPC, Raymond Yee, MD FRCPC FACC, George J. Klein, MD FRCPC FACC
Division of Cardiology, University of Western Ontario, London, Ontario, Canada 

Reprint requests: Lorne J. Gula, London Health Sciences Centre, Arrhythmia Service, 339 Windermere Road, London, Ontario CANADA, N6A 5A5.

Riassunto

Background

Verification of defibrillation efficacy by defibrillation threshold (DFT) testing during implantable cardioverter-defibrillator implantation is the current standard. Generally, defibrillation of ventricular fibrillation at 10 J below the maximum output of a device is felt to establish an adequate safety margin. Nonetheless, DFT testing adds to cost and carries some potential for morbidity, whereas its impact on outcomes in the modern era of defibrillator technology is unclear. We aimed to determine the frequency that DFT testing resulted in a change at device implant and to identify clinical and echocardiographic predictors of the need for DFT testing.

Methods

We reviewed all implantable cardioverter-defibrillators that were implanted at the London Health Sciences Centre (Ontario, Canada) from June 1999 to August 2003 and used multivariate analysis to determine variables associated with DFT test failures and elevated DFT values. When a defibrillation failure was not observed, a lowest energy to defibrillate (LED) was recorded.

Results

Among 168 implants, DFT testing was successful with a minimum 10-J safety margin in 152 (90%), whereas the remaining 16 required changes at device implant. In a multivariate analysis, use of amiodarone was independently associated with DFT failure (odds ratio, 4.6; 95% confidence interval, 1.2-17.0). Significantly higher mean DFT/LED values were observed among patients on amiodarone (1.36 J; P = .0041). Those with nonischemic cardiomyopathy had a higher mean DFT/LED compared with those with ischemic cardiomyopathy (1.44J; P = .028).

Conclusions

Use of amiodarone is associated with a 4-fold increase in risk of DFT failure and subsequent need for changes at implant to achieve a safe threshold. Defibrillation threshold testing appears to be most useful for patients taking amiodarone.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


© 2006  Mosby, Inc. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 152 - N° 6

P. 1104-1108 - dicembre 2006 Ritorno al numero
Articolo precedente Articolo precedente
  • Influence of age, sex, and atrial fibrillation recurrence on quality of life outcomes in a population of patients with new-onset atrial fibrillation: The Fibrillation Registry Assessing Costs, Therapies, Adverse events and Lifestyle (FRACTAL) study
  • Matthew R. Reynolds, Tara Lavelle, Vidal Essebag, David J. Cohen, Peter Zimetbaum
| Articolo seguente Articolo seguente
  • Plasminogen activator inhibitor type 1 gene polymorphisms and haplotypes are associated with plasma plasminogen activator inhibitor type 1 levels but not with myocardial infarction or stroke
  • Jingzhong Ding, Barbara J. Nicklas, Margaret D. Fallin, Nathalie de Rekeneire, Stephen B. Kritchevsky, Marco Pahor, Nicolas Rodondi, Rongling Li, Joseph M. Zmuda, Tamara B. Harris

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.