Electromagnetic interference with implantable cardiac pacemakers by video capsule - 18/08/11

Doi : 10.1016/S0016-5107(04)02834-2 
Sergio Dubner, MD, FACC, Yael Dubner, MD, Sebastian Gallino, MD, Liliana Spallone, MD, David Zagalsky, Horacio Rubio, MD, Joseph Zimmerman, MD, Eran Goldin, MD
Current affiliations: Department of Cardiology and Digestive Endoscopy, Clinica Suizo Argentina, Buenos Aires, Argentina, Department of Gastroenterology, Hadassah University Hospital, Jerusalem, Israel 

Reprint requests: Eran Goldin, MD, Head, Department of Gastroenterology, Hadassah University Hospital, Jerusalem, Israel.

Buenos Aires, Argentina

Abstract

Background

Electromagnetic fields, such as those generated by cellular phones and metal detectors, may interfere with normal pacemaker function. However, it remains unclear whether the wireless capsule endoscope interacts with implanted pacemakers. This prospective study evaluated potential interactions between the M2A video capsule (Given Imaging, Yoknam, Israel) and implanted pacemakers.

Methods

A total of 100 consecutive patients (70 men, 30 women) with an implanted pacemaker (95 on bipolar mode) were studied. The testing was performed with a functional testing device (Test Cap) for the Given Diagnostic System that reproduces the effect of the video capsule by transmitting at exactly the same frequency. During continuous electrocardiographic monitoring and recording, 100 tests were carried out without changing the pacemaker settings. Those with a positive result were retested 1 week later.

Results

The 100 pacemakers evaluated in the study population included the following: 70 dual chamber (11 DDD, 56 DDDR, 3 VDD) and 30 ventricular inhibited (12 VVI, 18 VVIR). In 4 of the 100 patients, pacemaker interference (noise-mode function forcing a synchronous mode) was registered during the Test Cap operation. Three patients had a dual-chamber pacemaker, and one had a single-chamber pacemaker. The interference was reproducible in all cases 1 week later. None of the implanted pacemakers tested was affected by oversensing.

Conclusions

Electromagnetic interferences with pacemakers from the M2A video capsule can occur, but this is without clinical significance. No potentially dangerous pacemaker inhibition was observed.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


© 2005  American Society for Gastrointestinal Endoscopy. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 61 - N° 2

P. 250-254 - febbraio 2005 Ritorno al numero
Articolo precedente Articolo precedente
  • Does the “Suspected Blood Indicator” improve the detection of bleeding lesions by capsule endoscopy?
  • Pierre-Nicolas D'Halluin, Michel Delvaux, Marie-Georges Lapalus, Sylvie Sacher-Huvelin, Emmanuel Ben Soussan, Laurent Heyries, Bernard Filoche, Jean-Christophe Saurin, Gerard Gay, Denis Heresbach
| Articolo seguente Articolo seguente
  • Capsule endoscopy vs. push enteroscopy and enteroclysis in suspected small-bowel Crohn's disease
  • André Kheng Ho Chong, Andrew Taylor, Ashley Miller, Oliver Hennessy, William Connell, Paul Desmond

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.