Endoscopic suturing without extracorporeal knots: a laboratory study - 18/08/11

Doi : 10.1016/S0016-5107(05)00331-7 
Bing Hu, MD, S.C. Sydney Chung, MD , Lawrence C.L. Sun, MPhil, James. Y.W. Lau, MD, Koichi Kawashima, BS, Tetsuya Yamamoto, BS, Peter B. Cotton, MD, Christopher J. Gostout, MD, Robert H. Hawes, MD, Anthony N. Kalloo, MD, Sergey V. Kantsevoy, MD, Pankaj J. Pasricha, MD
Current affiliations: Endoscopy Center, Prince of Wales Hospital, The Chinese University of Hong Kong, Hong Kong, China, Research and Development Department, Olympus Medical Systems Corp., Tokyo, Japan; The Apollo Group, Hong Kong, China (S.C.S.C.); Charleston, South Carolina (P.B.C., R.H.H.); Rochester, Minnesota (C.J.G.); Baltimore, Maryland (A.N.K., S.V.K.); Galveston, Texas (P.J.P.), USA 

Reprint requests: S. C. Sydney Chung, MD, Endoscopy Center, Prince of Wales Hospital, The Chinese University of Hong Kong, Shatin N.T., Hong Kong SAR, China.

Hong Kong, China

Abstract

Background

To eliminate cumbersome extracorporeal knotting, we designed a new endoscopic suturing device, the Eagle Claw V. The efficacy of the new device was tested on the Erlangen model and was compared with the initial extracorporeal knotting version (Eagle Claw II).

Methods

Segments of porcine splenic arteries were placed on the mucosal surface of the anterior wall of a pig stomach. The two ends of the artery were brought out through the gastric wall and were connected to a two-channel manometer. The Eagle Claw V has a curved needle with a detachable tip. After puncturing the tissue, the needle tip was engaged into a catching cartridge, where the suture had been embedded in a tightening mechanism. The suture could be tightened by simply pulling the end. The Eagle Claw V was used to plicate the splenic arteries mounted on the stomach. Suturing was considered secure if the suture could withstand endoluminal pressure greater than 200 mm Hg that lasted at least 10 seconds.

Results

Eleven of 15 stitches (73.3%) gained secure plication of the vessels. The suturing time (mean ± standard deviation 2.93 ± 0.80 minutes) was significantly faster than that of the Eagle Claw II (9.38 ± 1.51 minutes). The device consistently achieved penetration of the muscular layer, and 4 of 15 sutures included the serosa. The 4 failures were because of suture breakage in two, thread entanglement in one, and cartridge dislodgement in one.

Conclusions

The present prototype represents significant improvements in the ease of operation and the security of the suture, bringing the technique closer to clinical use for a variety of applications.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


 This study was an oral presentation at the annual Digestive Disease Week, May 15-20, 2004, New Orleans, Louisiana (Gastrointest Endosc 2004;59:AB114).


© 2005  American Society for Gastrointestinal Endoscopy. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 62 - N° 2

P. 230-233 - agosto 2005 Ritorno al numero
Articolo precedente Articolo precedente
  • Endoscopic variceal ligation safety throughout the GI tract
  • John S. Goff
| Articolo seguente Articolo seguente
  • Non-small-bowel lesions detected by capsule endoscopy in patients with obscure GI bleeding
  • Taya Kitiyakara, Warwick Selby

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.