Comparison of the Five-Electrode–Derived EASI Electrocardiogram to the Mason Likar Electrocardiogram in the Prehospital Setting - 21/08/11

Riassunto |
This study compared the 5-electrode-derived EASI electrocardiogram (ECG) with the conventional Mason-Likar ECG in 200 consecutive patients with chest pain transported to 3 hospitals by 2 different emergency medical services. No significant differences were observed between the 2 systems for the detection of relevant electrocardiographic abnormalities. A questionnaire administered to participating emergency medical personnel revealed a high degree of acceptability of the EASI ECG, with some participants commenting that the sternal and mid-axillary locations of the EASI electrodes made them easier to apply, especially to women, than conventional precordial electrodes.
Il testo completo di questo articolo è disponibile in PDF.| This study was supported by Philips Medical Systems, Oxnard, California, and ZOLL Medical Systems, Chelmsford, Massachusetts. |
Vol 96 - N° 3
P. 453-456 - agosto 2005 Ritorno al numeroBenvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.
Già abbonato a @@106933@@ rivista ?