Biometry and Formula Accuracy With Intraocular Lenses Used for Cataract Surgery in Extreme Hyperopia - 21/08/11

Doi : 10.1016/j.ajo.2007.02.043 
Robert E. MacLaren a, b, , Mythili Natkunarajah b, Yasmin Riaz a, Rupert R.A. Bourne a, Marie Restori a, Bruce D.S. Allan a, b
a Moorfields Eye Hospital, City Road, London, United Kingdom 
b Institute of Ophthalmology, University College London, London, United Kingdom. 

Inquiries to Robert E. MacLaren, Consultant Ophthalmologist, Moorfields Eye Hospital, 162 City Road, London EC1V 2PD, United Kingdom

Riassunto

Purpose

To audit intraocular lens (IOL) power predictions for cataract surgery in extreme hyperopia and to compare the accuracy across different biometry formulae and IOL types.

Design

A retrospective analysis of 76 eyes from 56 patients undergoing cataract surgery with IOLs ranging in power from 30 to 35 diopters (D).

Methods

Axial lengths, corneal powers and anterior chamber depths were measured with ultrasound or optical methods, and the IOLMaster (Carl Zeiss Meditech, Inc, Dublin, California, USA) software was used to predict the refractive outcome for each IOL used. Differences between the predicted and actual postoperative refraction were then analyzed for each formula.

Results

In practice, 55% of patients were within ±1.0 D of the refraction predicted by their surgeon. In theory, the Haigis formula would have given the smallest mean refractive error (+0.51 ± 0.12 D), followed by the Hoffer Q (−0.70 ± 0.14 D), Holladay 1 (−1.11 ± 0.13 D), and SRK/T formulae (−1.45 ± 0.14 D). The Haigis formula overpredicted the lens power required, which would have generated a myopic result. The other formulae underpredicted the lens power required and would have generated a hyperopic result. There was a significant difference between lens designs: the Haigis was more accurate for open-loop, whereas the Hoffer Q was more accurate for plate-haptic lenses. The anterior chamber depth measurement could also be used to predict changes in intraocular pressure after surgery.

Conclusion

This represents the largest published series to date of biometry predictions for cataract surgery in extreme hyperopia and confirms the Haigis formula to be the most accurate. A consistent difference between open-loop and plate-haptic lenses suggests that haptic design may influence the effective lens position in very small eyes. We further propose a simple formula to optimize the Haigis and Hoffer Q formulae in patients with extreme hyperopia.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


 Supplemental Material available at AJO.com.


© 2007  Elsevier Inc. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 143 - N° 6

P. 920 - giugno 2007 Ritorno al numero
Articolo precedente Articolo precedente
  • Long-term Effect of Optic Edge Design in a Silicone Intraocular Lens on Posterior Capsule Opacification
  • Wolf Buehl, Rupert Menapace, Oliver Findl, Thomas Neumayer, Matthias Bolz, Ana Prinz
| Articolo seguente Articolo seguente
  • Dietary Macronutrient Intake and Five-Year Incident Cataract: The Blue Mountains Eye Study
  • Bradley S. Townend, Marcia E. Townend, Victoria Flood, George Burlutsky, Elena Rochtchina, Jie Jin Wang, Paul Mitchell

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.