Reproducibility and Repeatability of Central Corneal Thickness Measurement in Keratoconus Using the Rotating Scheimpflug Camera and Ultrasound Pachymetry - 22/08/11

Doi : 10.1016/j.ajo.2007.07.021 
Ugo de Sanctis a, Alessandro Missolungi a, Bernardo Mutani a, , Lorenzo Richiardi b, Federico M. Grignolo a
a Department of Clinical Physiopathology, Ophthalmology Institute, University of Turin, Turin, Italy 
b Cancer Epidemiology Unit, CeRMS and CPO Piemonte, University of Turin, Turin, Italy. 

Inquiries to Bernardo Mutani, Department of Clinical Physiopathology, Ophthalmology Institute, University of Turin, Via Juvarra 19, Turin, Italy

Riassunto

Purpose

To assess repeatability, reproducibility, and agreement of rotating Scheimpflug camera (Pentacam Oculus, Wetzlar, Germany) and ultrasound pachymetry in measuring central thickness of keratoconic corneas.

Design

Method-comparison study.

Methods

In 33 patients with keratoconus (one eye per patient), two examiners each used both pachymetric methods to measure central corneal thickness (CCT); in the same session, measurements then were repeated by examiner 1 (A.M.). The difference between two examiners, and between first and second measurements by examiner 1, with both methods and the difference between the two pachymetric methods in measuring central thickness of keratoconic corneas were noted.

Results

With the rotating Scheimpflug camera, interexaminer correlation was higher (intra-class correlation coefficient [ICC], 0.98 vs 0.76) and inter-examiner variability was lower (95% limits of agreement [95% LoA], −14.8 to 13.8 μm vs −18.0 to +49.5 μm) than with ultrasound pachymetry. Both methods showed close first- to second-measurement correlation (ICC, > 90), but the rotating Scheimpflug camera had lower variability (95% LoA, −14.5 to 14.2 μm vs −27.4 to 26.0 μm). Mean CCT was 478.9 ± 34.6 μm with the rotating Scheimpflug camera and 486.6 ± 30 μm with ultrasound pachymetry. Although the mean difference was small (−7.8 μm), the 95% LoA (−43.8 to 28.2 μm) showed that the difference between the two methods can be considerable.

Conclusions

In keratoconic corneas, the rotating Scheimpflug camera provides measurements of central thickness that are more reproducible and repeatable than those obtained with ultrasound pachymetry. The rotating Scheimpflug camera seems to be suitable for disease staging and follow-up, when corneal thickness measurements may be repeated over time by different examiners.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


© 2007  Elsevier Inc. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 144 - N° 5

P. 712 - novembre 2007 Ritorno al numero
Articolo precedente Articolo precedente
  • Hepatitis C and Ocular Surface Disease
  • Christina Jacobi, Hartmut Wenkel, Arnd Jacobi, Klaus Korn, Claus Cursiefen, Friedrich E. Kruse
| Articolo seguente Articolo seguente
  • Phacoemulsification of Cataract in Patients Receiving Coumadin Therapy: Ocular and Hematologic Risk Assessment
  • Irina S. Barequet, Dan Sachs, Ayelet Priel, Yael Wasserzug, Uri Martinowitz, Joseph Moisseiev, Ophira Salomon

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.