A multicenter randomized comparison of the Endocapsule and the Pillcam SB - 23/08/11

Doi : 10.1016/j.gie.2007.12.037 
David R. Cave, MD, PhD , David E. Fleischer, MD, Jonathan A. Leighton, MD, Douglas O. Faigel, MD, Russell I. Heigh, MD, Virender K. Sharma, MD, Christopher J. Gostout, MD, Elizabeth Rajan, MD, Klaus Mergener, MD, Anne Foley, BS, Michael Lee, BS, Kanishka Bhattacharya, MD
Current affiliations: University of Massachusetts Memorial Medical Center (D.R.C., A.F., K.B.), Worcester, Massachusetts, Mayo Clinic (D.E.F., J.A.L., R.I.H., V.K.S., E.R., M.L.), Scottsdale, Arizona, Mayo Clinic (C.J.G.), Rochester, Minnesota, Oregon Health and Science University (D.E.F.), Portland, Oregon, Digestive Disease Specialists (K.M.), Tacoma, Washington, USA 

Reprint requests: David R. Cave, MD, PhD, University of Massachusetts Memorial Medical Center, 55 Lake Ave N, Worcester, MA 01655.

Worcester, Massachusetts, Scottsdale, Arizona, Rochester, Minnesota, Portland, Oregon, Tacoma, Washington, USA

Abstract

Background

Video capsule endoscopy has been shown to be the single most effective endoscopic procedure for identifying the source of obscure GI bleeding (OGIB).

Objective

Our purpose was to report on the Food and Drug Administration pivotal trial in which the Endocapsule (EC) (Olympus America, Allentown, Pa) was compared with the Pillcam SB (PSB) (Given Imaging, Yoqneam, Israel) in patients with OGIB.

Design

A novel trial design was used in which the EC and the PSB were swallowed by the same patient 40 minutes apart, in randomized order.

Setting

Four academic medical centers.

Patients

Patients with OGIB aged 18 to 85 years who had either been transfused or who had a hematocrit of <31% in males or <28% in females.

Interventions

Both video capsules were swallowed in random order. Videos from the PSB were read locally for patient management. All videos were then read by at least 2 independent readers for normal versus abnormal categorization, a diagnosis, capsule transit time, reading time, and a subjective assessment of image quality.

Main Outcome Measurement

Categorization of videos as either normal or abnormal.

Results

Data from 51 of 63 enrolled patients were analyzed. Nine patients were excluded for technical reasons and 3 for protocol violations. Twenty-four videos were read as normal and 14 as abnormal from both capsules. Disagreement occurred in 13. No adverse events were reported for either capsule. Overall agreement was 38 of 51 (74.5%) with a κ of 0.48, P = .008.

Limitations

Although ingestion order was randomized, the videos could not be read blind owing to a different shape of the image margin.

Conclusions

(1) Both devices were safe and had a comparable diagnostic yield within the range previously reported. (2) There was a subjective difference in image quality favoring the EC. (3) This study design provided unique information about capsule movement in the small intestine and the lack of electromechanical interference between 2 different capsules.

Il testo completo di questo articolo è disponibile in PDF.

Abbreviations : FDA, MST, OGIB, EC, PSB


Mappa


© 2008  American Society for Gastrointestinal Endoscopy. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 68 - N° 3

P. 487-494 - settembre 2008 Ritorno al numero
Articolo precedente Articolo precedente
  • Endoscopic balloon dilation of the sphincter of Oddi for stone extraction in the elderly: is the juice worth the squeeze?
  • James A. DiSario
| Articolo seguente Articolo seguente
  • Breaching the final frontier: the future of small-intestinal capsule endoscopy
  • Otto S. Lin

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.