EUS-guided radiofrequency ablation with a prototype electrode array system in an animal model (with video) - 23/08/11

Doi : 10.1016/j.gie.2009.03.008 
Shyam Varadarajulu, MD , Nirag C. Jhala, MD, Ernesto R. Drelichman, MD
Current affiliations: Division of Gastroenterology-Hepatology (S.V.), Department of Anatomic Pathology (N.C.J.), and Division of Gastrointestinal Surgery (E.R.D.), University of Alabama at Birmingham School of Medicine, Birmingham, Alabama 

Reprint requests: Shyam Varadarajulu, MD, University of Alabama at Birmingham Medical Center, 410 LHRB, 1530 3rd Avenue South, Birmingham, AL 35294.

Birmingham, Alabama, USA

Abstract

Background

Although previously reported in an animal model, the development of EUS-guided radiofrequency ablation (EUS-RFA) has been impeded because of a lack of a retractable needle electrode array that could safely and effectively ablate large areas.

Objective

To evaluate the feasibility and safety of performing EUS-RFA with a 19-gauge FNA needle fitted with an umbrella-shaped retractable needle electrode array.

Design and Setting

Endoscopic experimental study in a porcine survival model at a tertiary referral center animal laboratory.

Main Outcome Measures

Evaluate the safety and efficacy of the retractable needle electrode array for performing EUS-RFA.

Interventions

A 19-gauge EUS-FNA needle was modified and fitted with a retractable echogenic umbrella-shaped monopolar electrode array at its tip. The FNA needle was connected to a 200-W generator that has an impedance-based feedback system. EUS-RFA of the liver was attempted on 5 Yorkshire pigs. Although 1 pig was euthanized immediately after RFA to assess for immediate complications and pathological examination, the 4 others were kept alive for 7 days.

Results

At EUS, the needle electrode was well visualized and could be deployed in the liver without technical difficulty. During ablation, a round hyperechoic focus gradually surrounded the electrode tip. Tissue ablation was attained within 7 minutes, and the electrode array could be easily withdrawn into the needle assembly. The vital signs of all pigs remained stable throughout the procedure and until they were euthanized. Histopathology in all pigs revealed a discrete, well-demarcated spherical focus of complete coagulation necrosis measuring 2.6 cm in diameter and without damage to the surrounding liver parenchyma or vasculature.

Conclusions

In this experimental study, EUS-RFA of the liver was performed safely by using the retractable umbrella-shaped electrode array with effective coagulation necrosis of large areas.

Il testo completo di questo articolo è disponibile in PDF.

Abbreviations : RF, RFA


Mappa


 DISCLOSURE: The following authors disclosed financial relationships relevant to this publication: S. Varadarajulu: Speaker and consultant for Boston Scientific. E. R. Drelichman: Speaker and consultant for Boston Scientific. The other author disclosed no financial relationship relevant to this publication.
 If you want to chat with an author of this article, you may contact him at svaradarajulu@yahoo.com.


© 2009  American Society for Gastrointestinal Endoscopy. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 70 - N° 2

P. 372-376 - agosto 2009 Ritorno al numero
Articolo precedente Articolo precedente
  • Colonoscopic tips and tricks—advice from 3 master endoscopists
  • Tyler M. Berzin
| Articolo seguente Articolo seguente
  • Successful diagnostic and therapeutic intrauterine fetal interventions by natural orifice transluminal endoscopic surgery (with videos)
  • Samuel A. Giday, Jonathan M. Buscaglia, Janyne Althaus, Gianfranco Donatelli, Devi M. Krishnamurty, Xavier Dray, Dawn Ruben, David Liang, Ronald Wroblewski, Priscilla Magno, Eun Ji Shin, Anthony N. Kalloo

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.