Prospective trial comparing solid-state catheter and water-perfusion triple-lumen catheter for sphincter of Oddi manometry done at the time of ERCP - 23/08/11

Doi : 10.1016/j.gie.2008.11.044 
Peter V. Draganov, MD , Lukasz Kowalczyk, MD, Chris E. Forsmark, MD
Current affiliations: Division of Gastroenterology, Hepatology and Nutrition, University of Florida, Gainesville, Florida, USA 

Reprint requests: Peter V. Draganov MD, University of Florida, Div. Gastroenterology, Hepatology and Nutrition, 1600 SW Archer Rd, Room HD 602, PO Box 100214, Gainesville, FL 32610.

Gainesville, Florida, USA

Abstract

Background

Sphincter of Oddi manometry is the reference standard for the diagnosis of sphincter of Oddi dysfunction. Numerous studies have established ranges of normal values as well as typical readings in pathologic conditions. All these studies have been done using a water-perfused, triple-lumen catheter (TLC). A recently approved, new generation, solid-state catheter (SSC) has potential advantages, but concerns have been raised as to whether the pressures obtained by TLCs are reproducible by SSCs. To date, no data exist on the accuracy of sphincter of Oddi pressure measurements with the new-generation SSCs.

Objective

To evaluate the accuracy of the SSC by using the TLC as the reference standard.

Design

Prospective crossover trial.

Setting

A tertiary academic center.

Patients

Thirty patients with clinical indications for sphincter of Oddi manometry.

Interventions

Sphincter of Oddi manometry with TLC and SSC in the same patient.

Main Outcome Measurements

Accuracy of sphincter of Oddi pressure measurements.

Results

A total of 376 pressure measurements in 47 sphincter segments (24 biliary, 23 pancreatic) were obtained. Manometry results were abnormal in 10 of 24 biliary sphincters and 12 of 23 pancreatic sphincters. There was complete agreement on the final results of the sphincter of Oddi manometry (normal/abnormal) between the TLC and SSC (accuracy 100%). A split-plot analysis of the 378 individual measurements was performed. The P value of .9966 was insignificant, consistent with no catheter effect on the measurements.

Limitations

Lack of blinding.

Conclusions

Measurement of sphincter of Oddi pressures with the SSC is accurate, and results were essentially identical to those of the water-perfused catheter system.

Il testo completo di questo articolo è disponibile in PDF.

Abbreviations : FDA, SOD, SOM, SSC, TLC


Mappa


 DISCLOSURE: All authors disclosed no financial relationships relevant to this publication.
 If you want to chat with an author of this article, you may contact him at dragapv@medicine.ufl.edu.


© 2009  American Society for Gastrointestinal Endoscopy. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 70 - N° 1

P. 92-95 - luglio 2009 Ritorno al numero
Articolo precedente Articolo precedente
  • ERCP complication rates: how low can we go?
  • Michael E. Ryan
| Articolo seguente Articolo seguente
  • Individualizing colonoscopy screening by sex and race
  • Iris Lansdorp-Vogelaar, Marjolein van Ballegooijen, Ann G. Zauber, Rob Boer, Janneke Wilschut, Sidney J. Winawer, J. Dik F. Habbema

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.