Clinical performance of a specific algorithm to reconfirm self-terminating ventricular arrhythmias in current implantable cardioverter-defibrillators - 03/09/11

Doi : 10.1016/S0002-9149(01)01844-6 
Tanja M Raedle-Hurst, MD a, , Johannes Wiecha, MD a, Joerg O Schwab, MD a, Heiko Schmitt, MD a, Michael Hinrichs a, Harald Tillmanns, MD a, Bernd Waldecker, MD a
a Department of Cardiology, University of Giessen, Giessen, Germany 

*Address for reprints: Tanja M. Raedle-Hurst, MD, Department of Cardiology, University of Giessen, Klinikstrasse 36, 35385 Giessen, Germany

Abstract

Inappropriate shock therapy is a frequent problem in patients with implantable cardioverter-defibrillators (ICDs), caused mostly by supraventricular rhythms. Self-terminating ventricular arrhythmias (STVAs), however, may also lead to inappropriate shock discharges even in ICDs with abortive shock capabilities. The aim of this study was to evaluate the clinical performance of a specific ventricular tachycardia/ventricular fibrillation (VT/VF) reconfirmation algorithm implemented in current ICD devices from Medtronic to prevent inappropriate shock discharges due to STVAs. A total of 161 STVA episodes were documented in 59 of 150 patients (39%) within a mean follow-up of 30 ± 20 months and resulted in 25 inappropriate shock discharges in 15 of 150 patients (10%) despite activation of the reconfirmation algorithm. The first synchronization interval of the algorithm was met in 92% of STVA episodes with and even 38% of STVA episodes without shock delivery. A reduced incidence of inappropriate shocks due to STVAs was found with tachycardia/fibrillation detection intervals (TDI/FDI) programmed to shorter cycle lengths ≤280 ms or the use of the first 2 cycles after the end of charging to be considered for reconfirmation only. Thus, inappropriate shocks due to STVAs still occur in 10% of patients with ICDs despite activation of a specific VT/VF reconfirmation algorithm, and are mainly caused by meeting the first synchronization interval that therefore should be shortened in cycle length. Moreover, to reduce the likelihood of inappropriate shocks, the VF reconfirmation algorithm should be optimized by basing the synchronization intervals exclusively on the FDI with short cycle lengths or using the first 2 cycles for reconfirmation only.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


© 2001  Excerpta Medica Inc. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 88 - N° 7

P. 744-749 - ottobre 2001 Ritorno al numero
Articolo precedente Articolo precedente
  • Efficacy of cholesterol levels and ratios in predicting future coronary heart disease in a Chinese population
  • Tzung-Dau Wang, Wen-Jone Chen, Kuo-Liong Chien, Sophia Seh-Yi Su, Hsiu-Ching Hsu, Ming-Fong Chen, Chiau-Suong Liau, Yuan-Teh Lee
| Articolo seguente Articolo seguente
  • Long-term infection rates associated with the pectoral versus abdominal approach to cardioverter- defibrillator implants
  • Theofanie Mela, Brian A McGovern, Hasan Garan, Gus J Vlahakes, David F Torchiana, Jeremy Ruskin, Joseph M Galvin

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.