PRESSURE-CONTROLLED AND VOLUME-CYCLED MECHANICAL VENTILATION - 11/09/11

Doi : 10.1016/S0272-5231(05)70323-3 
Andrew W. McKibben, MD, Sue A. Ravenscraft, MD
a Department of Pulmonary and Critical Care Medicine, St. Paul-Ramsey Medical Center, St. Paul, (AWM); Park Nicollet Clinic; and Department of Medicine, University of Minnesota Medical School, Minneapolis (SAR), Minnesota 

Riassunto

Pressure and volume modes of mechanical ventilation have been available for many years but not as options on the same ventilator until the early 1980s.21 The clinician now is faced with many choices when managing a mechanically ventilated patient. This article provides basic information on flow-limited, volume-cycled ventilation (VCV) and pressure-controlled ventilation (PCV) so comparisons and contrasts can be made between the two modes. The authors also review recommendations and potential problems commonly encountered when converting from one mode to the other. A good working understanding of volume and pressure modes of ventilation should allow the clinician to transfer from one mode to the other without difficulty. The final section of this article reviews some of the pertinent clinical studies and recent direct comparisons of VCV and PCV.

Although many of the principles discussed pertain to positive-pressure ventilation in general, this article focuses on assist-control VCV and PCV (pressure preset). It is flow that is controlled and cycled in VCV, but for the sake of simplicity, the authors refer to it as VCV. Aside from PCV, pressure modes of ventilation include pressure support and airway pressure release ventilation.24, 73 In addition, several options exist for combination modes (i.e., synchronized intermittent mandatory ventilation with pressure support) as well as several newer ventilatory modes available on the latest generation of mechanical ventilators (i.e., pressure-regulated volume control, volume-assured pressure support, and volume support). These modes attempt to combine the most desirable features of volume-based and pressure-based ventilation.40

The primary difference between PCV and the more commonly used VCV is that clinician-set inputs and measured outputs are exchanged. PCV does not deliver a predetermined tidal volume (VT) but applies and maintains a set airway pressure throughout inspiration for a prescribed length of time. VT is set in VCV and pressures are measured; in PCV, pressure is set and volume is measured. The measured outputs in both modes are governed by the impedance characteristics of the respiratory system (resistance and compliance) and the ventilator settings (i.e., frequency, inspiratory time).

Il testo completo di questo articolo è disponibile in PDF.

Mappa


 Address reprint requests to Sue A. Ravenscraft, MD, Park Nicollet Clinic, 3800 Park Nicollet Boulevard, St. Louis Park, MN 55416


© 1996  W. B. Saunders Company. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.© 1992  © 1992  © 1995  © 1992  © 1995 
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 17 - N° 3

P. 395-410 - settembre 1996 Ritorno al numero
Articolo precedente Articolo precedente
  • AUTO–POSITIVE END-EXPIRATORY PRESSURE AND DYNAMIC HYPERINFLATION
  • V. Marco Ranieri, Salvatore Grasso, Tommaso Fiore, Rocco Giuliani
| Articolo seguente Articolo seguente
  • NEW MODES OF MECHANICAL VENTILATION
  • Neil R. MacIntyre

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.