Abbonarsi

New laser soldering-based closures: a promising method in natural orifice transluminal endoscopic surgery - 20/06/12

Doi : 10.1016/j.gie.2012.03.006 
Serge Bogni, PhD 1, Maria-Anna Ortner, MD 2, Istvan Vajtai, MD 3, Christian Jost, MD 2, Michael Reinert, MD 4, Bernard Dallemagne, MD 5, Martin Frenz, PhD 1,
1 Department of Biomedical Photonics, Institute of Applied Physics, University of Bern, Bern, Switzerland 
2 Department of Visceral Surgery and Medicine, University Hospital Inselspital, Bern, Switzerland 
3 Institute of Pathology, University Hospital Inselspital, Bern, Switzerland 
4 Department of Neurosurgery, University Hospital Inselspital, Bern, Switzerland 
5 Institut de Recherche contre les Cancers de l'Appareil Digestif (IRCAD) and Department of Digestive and Endocrine Surgery, Les Hôpitaux Universitaires de Strasbourg, Strasbourg, France 

Reprint requests: Martin Frenz, PhD, Department of Biomedical Photonics, Institute of Applied Physics, University of Bern Switzerland, Sidlerstrasse 5, 3012 Bern, Switzerland

Riassunto

Background

Complete closure of gastrotomy is the linchpin of safe natural orifice transgastric endoscopic surgery.

Objective

To evaluate feasibility and efficacy of a new method of gastrotomy closure by using a sutureless laser tissue-soldering (LTS) technique in an ex vivo porcine stomach.

Design

In vitro experiment.

Setting

Experimental laboratory.

Interventions

Histological analysis and internal and external liquid pressure with and without hydrochloric acid exposure were determined comparing gastrotomy closure with LTS and with hand-sewn surgical sutures.

Main Outcome Measurements

Comparison of LTS and hand-sewn surgical gastrotomy closure. The primary outcome parameter was the internal leak pressure. Secondary parameters were the difference between internal and external leak pressures, the impact of an acid environment on the device, histological changes, and feasibility of endoscopic placement.

Results

The internal liquid leak pressure after LTS was almost twice as high as after hand-sewn surgical closure (416 ± 53 mm Hg vs 229 ± 99 mm Hg; P = .01). The internal leak pressure (416 ± 53 mm Hg) after LTS was higher than the external leak pressure (154 ± 46 mm Hg; P < .0001). An acidic environment did not affect leak pressure after LTS. Endoscopic LTS closure was feasible in all experiments. Histopathology revealed only slight alterations beneath the soldering plug.

Limitations

In vitro experiments.

Conclusions

Leak pressure after LTS closure of gastrotomy is higher than after hand-sewn surgical closure. LTS is a promising technique for closure of gastrotomies and iatrogenic perforations. Further experiments, in particular survival studies, are mandatory.

Il testo completo di questo articolo è disponibile in PDF.

Abbreviations : EASIE, LTS, NOTES, PCL


Mappa


 DISCLOSURE: The authors disclosed no financial relationships relevant to this publication. Supported by Swiss Commission of the Encouragement of Scientific Research contract grant 11478.1pfls-ls.
 If you would like to chat with an author of this article, you may contact Dr Frenz at frenz@iap.unibe.ch.


© 2012  American Society for Gastrointestinal Endoscopy. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 76 - N° 1

P. 151-158 - luglio 2012 Ritorno al numero
Articolo precedente Articolo precedente
  • Construct validation of a physical model colonoscopy simulator
  • Annaliese M. Plooy, Andrew Hill, Mark S. Horswill, Alanna St. G. Cresp, Marcus O. Watson, Soong-Yuan Ooi, Stephan Riek, Guy M. Wallis, Robin Burgess-Limerick, David G. Hewett
| Articolo seguente Articolo seguente
  • Continuing Medical Education Exam: July 2012
  • Raquel E. Davila, Jeffrey H. Lee, William Ross, Shou-Jiang Tang, G.S. Raju, Glenn M. Eisen

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.