Abbonarsi

Poor short-term outcomes after computer-assisted rotating-platform total knee arthroplasty with a deep-trochlear-groove femoral component: Analysis of 19 patients - 30/03/13

Doi : 10.1016/j.otsr.2012.07.010 
A. Akakpo a, , b , G. Dereudre a, b, N. Fouilleron a, b, G. Pasquier a, b, H. Migaud a, b
a Nothern France Lille University, 59000 Lille, France 
b Department of Orthopaedics, Salengro Hospital, Lille Regional Teaching Hospital Center, place de Verdun, 59037 Lille cedex, France 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Summary

Background

Congruent rotating tibial plateaus are designed to minimise wear after total knee arthroplasty (TKA). The Score™ prosthesis has a congruent rotating tibial plateau, a deep trochlear groove, and uses a computer-assisted navigation system for ligament balancing. Although this prosthesis is widely utilized, no accurate data on outcomes are available.

Hypothesis

The innovative features of the Score™ prosthesis, most notably patellar replacement with a highly constrained femoral component, do not jeopardize implant survival.

Patients and methods

In a pilot study, we retrospectively evaluated outcomes of 19 patients treated with Score™ knee replacement between February and October 2006 (mean age, 66.8years; range, 58–82years). The evaluation criteria were the International Knee Society (IKS) scores and prosthesis survival rate estimated using Kaplan-Meier plots with failure defined as revision need to change the prosthesis.

Results

Mean follow-up was 35.3months. The IKS knee score increased from 27.4 (5–60) preoperatively to 81.4 (45–99) at last follow-up (P<0.0001). Mean mechanical axis was 181.2° (180–186°), with 16 between 180° and 183°. Revision surgery was required in five cases (for patellar complications with combined motion-range limitation in flexion (<90°) and extension (5–20°) in three cases, isolated motion-range limitation in one case, and recurvatum deformity with instability in one case). Prosthesis survival was 82% (73–91%) after 24months and 65% (51–78%) after 44months.

Discussion

The deep trochlear groove femoral component resulted in patellar complications, which were the most common reasons for revision surgery, together with motion-range limitation and instability possibly related to improper use of the navigation system. This small retrospective case-series study showed an unusually low prosthesis survival rate probably related to the implant design. We no longer use the Score™ prosthesis, despite the availability of a dedicated navigation system, and we recommend careful monitoring of patients who have this prosthesis.

Level of evidence

Level IV, retrospective study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Knee prosthesis, Rotating platform, Computer-assisted implant surgery, Patello-femoral, Instability


Mappa


© 2012  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 99 - N° 2

P. 183-190 - aprile 2013 Ritorno al numero
Articolo precedente Articolo precedente
  • Unstable pelvic ring injury with hemodynamic instability: What seems the best procedure choice and sequence in the initial management?
  • S. Abrassart, R. Stern, R. Peter
| Articolo seguente Articolo seguente
  • Simultaneous bilateral total knee arthroplasty. A multicenter feasibility study
  • J.-Y. Jenny, C. Trojani, J.-L. Prudhon, C. Vielpeau, D. Saragaglia, C. Houillon, T. Ameline, F. Steffan, B. Bugnas, J. Arndt, Hip and Knee Surgery French Society (SFHG)

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.