Correlation Between Corneal and Scleral Pneumatonometry: An Alternative Method for Intraocular Pressure Measurement - 23/10/13

Doi : 10.1016/j.ajo.2013.05.045 
Michael A. Kapamajian, Jose de la Cruz, Joelle A. Hallak, Thasarat S. Vajaranant
 Illinois Eye and Ear Infirmary, University of Illinois at Chicago, Department of Ophthalmology and Visual Sciences, Chicago, Illinois 

Inquiries to Thasarat S. Vajaranant, Department of Ophthalmology and Visual Sciences, 1855 W Taylor St, Chicago, IL 60612

Abstract

Purpose

To evaluate scleral pneumatonometry as an alternative method for measuring intraocular pressure (IOP).

Design

Prospective cross-sectional study.

Methods

Adult subjects with healthy eyes were recruited from the Comprehensive Eye Service at the University of Illinois Eye and Ear Infirmary from August 2008 through February 2009. Study measurements included corneal pneumatonometry (IOPk), scleral pneumatonometry (IOPs), axial length (AL), spherical equivalent (SE), and central corneal thickness (CCT). Main outcome measures were scleral IOP and corneal IOP.

Results

Analysis included a monocular data set from single eyes of 97 subjects (age: 18-82 years). IOPs was consistently higher than IOPk, and correlated positively with IOPk (r = 0.57, P < .0001), age (r = 0.51, P < .0001), and SE (r = 0.32, P = .0002). The difference between scleral and corneal IOP (IOPs − IOPk) correlated positively with IOPs (r = 0.89, P < .0001), age (r = 0.57, P < .0001), and SE (r = 0.34, P < 0.0001). Bland-Altman analysis for agreement between scleral and corneal pneumatonometry measurements showed a mean difference of 8.08 mm Hg, with the 95% limit of agreement between -3.47 and 19.64 mm Hg. Regression analysis yielded the following equation: IOPk = 11.9 + 0.32(IOPs) − 0.05(Age).

Conclusions

Scleral pneumatonometry correlates positively with corneal pneumatonometry and is more accurate at lower values and in younger patients. When adjusted for age, scleral pneumatonometry may be an adequate alternative in situations where corneal measurements are impractical.

Il testo completo di questo articolo è disponibile in PDF.

Mappa


 Dr Kapamajian is currently a practicing ophthalmologist at the California Eye and Ear Specialists, San Gabriel, California.


© 2013  Elsevier Inc. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 156 - N° 5

P. 902 - novembre 2013 Ritorno al numero
Articolo precedente Articolo precedente
  • Anterior Segment Optical Coherence Tomography of Long-Term Phakic Angle-Supported Intraocular Lenses
  • Jorge L. Alio, Alessandro Abbouda, Pablo Peña-Garcia
| Articolo seguente Articolo seguente
  • Macular Structure Parameters as an Automated Indicator of Paracentral Scotoma in Early Glaucoma
  • Yugo Kimura, Masanori Hangai, Akiko Matsumoto, Tadamichi Akagi, Hanako O. Ikeda, Shinji Ohkubo, Kazuhisa Sugiyama, Aiko Iwase, Makoto Araie, Nagahisa Yoshimura

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.