Hepatic and Renal Function with Successful Long-term Support on a Continuous Flow Left Ventricular Assist Device - 25/02/14

Riassunto |
Introduction |
Data regarding the long-term clinical effects of a continuous flow left ventricular assist device (CF-LVAD) on hepato-renal function is limited. Hence our aim was to assess changes in hepato-renal function over a one-year period in patients supported on a CF-LVAD.
Methods |
During the study period 126 patients underwent CF-LVAD implant. Changes in hepato-renal laboratory parameters were studied in 61/126 patients successfully supported on a CF-LVAD for period of one year. A separate cohort of a high-risk group (HCrB) of patients (56/126) with a serum creat>1.9mg/dL (168μmol/L) (75th percentile) or a serum bil>1.5mg/dL (25.65μmol/L) (75th percentile) was created. Changes in serum creatinine and bilirubin were analysed at regular intervals for this group along with the need for renal replacement therapy.
Results |
Baseline creatinine and blood urea nitrogen (BUN) for the entire cohort was 1.4[1.2,1.9mg/dL] [123.7(106,168)μmol/L) and 27[20,39.5mg/dL] [9.6(7.1,14.1)mmol/L] respectively. After an initial reduction at the end of one month [1(0.8,1.2)mg/dL; 88(70,105)μmol/L] (p<0.0001), a gradual increase was noted over the study period to reach (1.25[1.1,1.5]mg/dL; 106(97.2,132.6)μmol/L] (p=0.0003). The serum bilirubin normalised from a [1(0.7,1.55)mg/dL] [17(18.8,25.7)μmol/L) to 0.9(0.6,1.2)mg/dL [15.4(10.2,20.5) μmol/L] (p=0.0005) and continued to decline over one year. Improvement in the synthetic function of the liver was demonstrated by a rise in the serum albumin levels to reach 4.3[4.1,4.5] [43(41,45)gm/L] at the end of one year (p<0.0001).
The baseline serum creatinine and bilirubin for the high-risk cohort (HCrB) was 1.9(1.3,2.4)mg/dL [168(115,212)μmol/L] and 1.7(1.00,2.4)mg/dL [29(17.1,68.4)μmol/L] respectively. The high-risk cohort (HCrB) demonstrated a trend towards higher 30-day mortality (p=0.06). While the need for temporary renal replacement therapy was higher in this cohort (16% vs. 4%; p=0.03), only 3% need it permanently. A significant reduction in creatinine was apparent at the end of one month [1.1(0.8,1.4)mg/dL; 97(70.7,123.7)μmol/L] (p<0.0001) and then remained stable at [1.3(1.1,1.5)mg/dL; 115(97,132.6)μmol/L]. Bilirubin demonstrated a 30% decline over one month and then remained low at [0.7(0.5,0.8)mg/dL; 62(44,70)μmol/L] p=0.0005 compared to the pre-operative baseline.
Conclusion |
Hepato-renal function demonstrates early improvement and then remains stable in the majority of patients on continuous flow left ventricular assist device support for one year. High-risk patients demonstrate a higher 30-day mortality and temporary need for renal replacement therapy. Yet even in this cohort, improvement is present over a period of one year on the device, with a minimal need for permanent haemodialysis.