Computerized 3-dimensional localization of a video capsule in the abdominal cavity: validation by digital radiography - 15/03/14

Doi : 10.1016/j.gie.2013.11.022 
Neil Marya, MD 1, , Andrew Karellas, PhD 1, Anne Foley, BA, CCRC 1, Abhijit Roychowdhury, MD 1, 2, David Cave, MD, PhD 1, 2
1 University of Massachusetts Medical School, Worcester, Massachusetts, USA 
2 UMass Memorial Healthcare, Worcester, Massachusetts, USA 

Reprint requests: Neil Marya, MD, University of Massachusetts Medical School, Worcester, MA 01532.

Abstract

Background

Wireless video capsule endoscopy allows the noninvasive visualization of the small intestine. Currently, capsules do not provide localization information while traversing the GI tract.

Objective

To report on the radiological validation of 3-dimensional localization software incorporated in a newly developed capsule. By using radiofrequency transmission, the software measures the strength of the capsule's signal to locate the position of the capsule.

Setting

This study was performed at the University of Massachusetts Medical Center, Worcester, Mass.

Patients

Thirty healthy volunteers consented to the experimental procedure.

Design

After ingestion of the capsule, subjects had 5 sets of anteroposterior and lateral radiographs taken every 30 minutes while the software calculated the position of the capsule. By using the radiographs, we calculated the location of the capsule in the abdominal cavity and compared the results with those generated by the software.

Results

Average error (and standard deviation) among the 3-dimensional coordinates was X, 2.00 cm (1.64); Y, 2.64 cm (2.39); and Z, 2.51 cm (1.83). The average total spatial error among all measurements was 13.26 cm3 (22.72). There was a correlation between increased subject body mass index and the 3-dimensional software measurement error.

Limitations

This study was performed in healthy volunteers and needs further validation in patients with small intestinal disorders.

Conclusions

The new 3-dimensional software provides localization of the capsule consistent with radiological observations. However, further validation of the software's clinical utility is required with a prospective clinical trial.

Il testo completo di questo articolo è disponibile in PDF.

Abbreviations : AP, BMI, RF


Mappa


 DISCLOSURE: The following author disclosed financial relationships relevant to this publication: Dr Cave is a consultant to Olympus Medical Systems and Capso Vision. All other authors disclosed no financial relationships relevant to this publication. This study was supported by Olympus, Tokyo, Japan.
 If you would like to chat with an author of this article, you may contact Dr Marya at neil.marya@umassmemorial.org.


© 2014  American Society for Gastrointestinal Endoscopy. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 79 - N° 4

P. 669-674 - aprile 2014 Ritorno al numero
Articolo precedente Articolo precedente
  • Near-infrared fluorescence cholangiopancreatoscopy: initial clinical feasibility results
  • Jürgen Glatz, P. Beatriz Garcia-Allende, Valentin Becker, Maximilian Koch, Alexander Meining, Vasilis Ntziachristos
| Articolo seguente Articolo seguente
  • GI bleeding and obstructive jaundice caused by osteoclast-like giant cell tumor of the head of the pancreas eroding into the duodenum
  • Maryconi M. Jaurigue, Palaniappan Manickam, Mitual Amin, Mitchell S. Cappell

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
L'accesso al testo integrale di questo articolo richiede un abbonamento.

Già abbonato a @@106933@@ rivista ?

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.