Abbonarsi

Post-breakfast closed-loop glucose control is improved when accompanied with carbohydrate-matching bolus compared to weight-dependent bolus - 17/06/14

Doi : 10.1016/j.diabet.2013.12.001 
A. Haidar a, , g, 1 , D. Farid a, g, 1, A. St-Yves a, 1, V. Messier a, V. Chen b, D. Xing b, A.-S. Brazeau a, C. Duval a, B. Boulet c, L. Legault d, R. Rabasa-Lhoret a, e, f
a Institut de Recherches Cliniques de Montréal, Montreal, Canada 
b Jaeb Center for Health Research, Tampa, FL, USA 
c Centre for Intelligent Machines, McGill University, Montreal, Canada 
d Montreal Children's Hospital, McGill University Health Centre, Montreal, Canada 
e Nutrition department, Faculty of medicine, Université de Montréal, Montreal, Canada 
f Montreal diabetes research center, Montreal, Canada 
g Experimental Medicine, McGill University, Montreal, Canada 

Corresponding author. Institut de Recherches Cliniques de Montréal, 110, des Pins West, H2W 1R7, Montreal, Canada. Tel.: +514 987 5500x3238; fax: +514 987 5670.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Aim

We compared post-breakfast closed-loop glucose control either matched with a carbohydrate-matching bolus or a weight-dependent bolus.

Methods

Twelve adults with type 1 diabetes consumed a 75g CHO breakfast on two occasions. In random order, the breakfast was accompanied by a full carbohydrate-matching insulin bolus (8.30U [7.50U–11.50U]) or a partial weight-dependent insulin bolus (0.047U/kg; 3.45U [2.95U–3.75U]). Postprandial glucose was regulated by sensor-responsive insulin and glucagon delivery.

Results

Glucose control after the weight-dependent bolus was safe and feasible (glucose values returned to pre-prandial levels after 5h). However, 5-hr incremental area under the curve and percentage of time above 10mmol/L were lower after the full bolus compared to the partial bolus (IAUC, 2.1 [0.8–4.2]mmol/L/hr vs 8.3 [6.5–11.4] mmol/L/hr; time in hyperglycaemia, 24% [6%–29%] vs 50% [25%–63%]; P<0.001).

Conclusions

Post-breakfast closed-loop glucose control without carbohydrate counting, but based on weight-dependent bolus is feasible but a carbohydrate-matching bolus provides better glucose control.

Clinical trial registry

NCT01519102

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Type 1 diabetes, Closed-loop delivery, Postprandial control, Glucagon, Artificial pancreas


Mappa


© 2014  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 40 - N° 3

P. 211-214 - giugno 2014 Ritorno al numero
Articolo precedente Articolo precedente
  • Body weight, weight gain and hyperglycaemia are associated with hypertensive disorders of pregnancy in women with gestational diabetes
  • B. Barquiel, L. Herranz, C. Grande, I. Castro-Dufourny, M. Llaro, P. Parra, M.A. Burgos, L.F. Pallardo
| Articolo seguente Articolo seguente
  • Effects of adipose tissue distribution on maximum lipid oxidation rate during exercise in normal-weight women
  • L. Isacco, D. Thivel, M. Duclos, J. Aucouturier, N. Boisseau

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.