Abbonarsi

Does the PFNA™ nail limit impaction in unstable intertrochanteric femoral fracture? A 115 case-control series - 07/02/15

Doi : 10.1016/j.otsr.2014.11.009 
M. Hélin a, b, A. Pelissier a, b, P. Boyer a, b, T. Delory c, C. Estellat c, P. Massin a, b,
a Département de chirurgie orthopédique, hôpitaux universitaires Paris Nord Val-de-Seine, site Bichat, 46, rue Henri-Huchard, 75877 Paris cedex 18, France 
b Université Paris Diderot, Sorbonne Paris Cité, EA REMES, 75010 Paris, France 
c Département de biostatistiques, hôpitaux universitaires Paris Nord Val-de-Seine, site Bichat, 46, rue Henri-Huchard, 75877 Paris cedex 18, France 

Corresponding author. Département de chirurgie orthopédique, hôpitaux universitaires Paris Nord Val-de-Seine, site Bichat, 46, rue Henri-Huchard, 75877 Paris cedex 18, France. Tel.: +33 1 40 87 52 23; fax: +33 1 40 87 52 92.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Introduction

Femoral neck shortening after dynamic fixation of extra-capsular fracture may impair functional results, but is rarely assessed. The present study measured impaction in stable and unstable fractures (on the Ender classification) treated by PFNA™ nail. The objectives were: 1) to validate the Ender classification to assess fracture stability; 2) to determine whether neck shortening and head purchase quality varied with stability; and 3) to determine the functional impact of femoral neck shortening.

Hypothesis

The study hypothesis was that the PFNA™ nail stabilizes unstable as well as stable fractures.

Materials and Methods

One hundred and fifteen consecutive patients, aged over 70 years, operated on for intertrochanteric fracture using the PFNA™ nail were followed up prospectively for 6 months. Multivariate analysis, including age, gender, assembly quality and body-mass index, was applied to assess the predictive power of the Ender classification with respect to femoral neck shortening. Secondly, patients were grouped according to stable versus unstable fracture (n=70 and 45, respectively), and impaction and femoral head purchase were assessed on a dedicated radiographic protocol. Functional results were assessed on Parker score.

Results

In the unstable fracture group, 3 assembly failures required revision by total hip replacement. Ender grade>2 was significantly predictive of>5mm neck shortening. Neck shortening was greater in unstable fracture: 8.1±8.4mm (range, 4–32mm), versus 2.5±3.7mm (range, 3–14mm) (P=0.0004). Mean blade cut-through was 1.2±2.9mm (range, 1–12mm) in unstable fracture, versus 0.3±1.3mm (range, 1–6mm) (P=0.02). Mean cut-out was 2.3±6mm (range, 2–21mm) in unstable fracture, versus 0.5±2.6mm (range, 1–8mm) (P=0.03). Parker scores diminished comparably in the two groups, without significant difference at follow-up: 3.9±2.6 (range, 0–9) in stable and 3.1±1.9 (range, 0–8) in unstable fracture; reduction in Parker score showed no correlation with femoral neck shortening (r=0.013, P=0.88).

Discussion

The PFNA™ nail provides poorer stabilization of unstable compared to stable fracture. Femoral neck shortening should be taken into account in assessing internal fixation hardware perfomances.

Level of evidence

Level III. Prospective case-control study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Intertrochanteric fracture, Internal fixation, O.R.I.F unstable fracture, Secondary displacement, Fixation failure


Mappa


© 2014  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 101 - N° 1

P. 45-49 - febbraio 2015 Ritorno al numero
Articolo precedente Articolo precedente
  • Effectiveness and safety of the posterior approach with soft tissue repair for primary total hip arthroplasty: A meta-analysis
  • D. Zhang, L. Chen, K. Peng, F. Xing, H. Wang, Z. Xiang
| Articolo seguente Articolo seguente
  • Evolution of slipped capital femoral epiphysis after in situ screw fixation at a mean 11 years’ follow-up: A 222 case series
  • E. Nectoux, J. Décaudain, F. Accadbled, A. Hamel, N. Bonin, P. Gicquel, French society of Orthopedic, Traumatologic Surgery (SoFCOT) f

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.