Dose reduction with iterative reconstruction: Optimization of CT protocols in clinical practice - 07/05/15

Doi : 10.1016/j.diii.2015.02.007 
J. Greffier a, , F. Macri a, A. Larbi a, A. Fernandez a, E. Khasanova a, b, F. Pereira a, C. Mekkaoui a, c, J.P. Beregi a
a Department of Radiology, University Hospital Center of Nîmes, EA 2415, Bd Prof Robert-Debré, 30029 Nîmes cedex, France 
b Dana-Farber Cancer Institute, Harvard Medical School, Boston, USA 
c Harvard Medical School, Massachusetts General Hospital, Department of Radiology, Martinos Center for Biomedical Imaging, Boston, USA 

Corresponding author. CHU de Nîmes, Bd Prof Robert-Debré, 30029 Nîmes Cedex 9, France.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Objectives

To create an adaptable and global approach for optimizing MDCT protocols by evaluating the influence of acquisition parameters and Iterative Reconstruction (IR) on dose reduction and image quality.

Materials and methods

MDCT acquisitions were performed on quality image phantom by varying kVp, mAs, and pitch for the same collimation. The raw data were reconstructed by FBP and Sinogram Affirmed Iterative Reconstruction (SAFIRE) with different reconstruction kernel and thickness. A total of 4032 combinations of parameters were obtained. Indices of quality image (image noise, NCT, CNR, SNR, NPS and MTF) were analyzed. We developed a software in order to facilitate the optimization between dose reduction and image quality. Its outcomes were verified on an adult anthropomorphic phantom.

Results

Dose reduction resulted in the increase of image noise and the decrease of SNR and CNR. The use of IR improved these indices for the same dose without affecting NCT and MTF. The image validation was performed by the anthropomorphic phantom. The software proposed combinations of parameters to reduce doses while keeping indices of the image quality adequate. We observed a CTDIvol reduction between −44% and −83% as compared to the French diagnostic reference levels (DRL) for different anatomical localization.

Conclusion

The software developed in this study may help radiologists in selecting adequate combinations of parameters that allows to obtain an appropriate image with dose reduction.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Multidetector CT, Iterative reconstruction, Dose reduction, Image quality, Patient safety

Abbreviations : CNR, CTDI, DRL, FBP, FOV, IR, LDPE, LSF, MDCT, MTF, NCT, NPS, PSF, ROI, SAFIRE, SNR, VBA


Mappa


© 2015  Éditions françaises de radiologie. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 96 - N° 5

P. 477-486 - maggio 2015 Ritorno al numero
Articolo precedente Articolo precedente
  • Analysis of subsequent publication of scientific orally presented abstracts of the French National Congress of Radiology. Part II: Focus on the French abstracts
  • V. Dangouloff-Ros, M. Ronot, M. Lagadec, V. Vilgrain
| Articolo seguente Articolo seguente
  • CT pulmonary angiogram with 60% dose reduction: Influence of iterative reconstructions on image quality
  • F. Pontana, D. Moureau, B. Schmidt, A. Duhamel, J.-B. Faivre, K. Yasunaga, J. Remy, M. Remy-Jardin

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.