Abbonarsi

Poor reproducibility of the MRI measurement of distal femoral torsion - 08/12/15

Doi : 10.1016/j.otsr.2015.09.028 
M. Ollivier a, b, , C. Stelzlen a, P. Boisrenoult a, N. Pujol a, P. Beaufils a
a Département de chirurgie orthopédique, centre hospitalier de Versailles, 78150 Le Chesnay, France 
b Aix-Marseille université, CNRS, ISM UMR 7287, 13284 Marseille, France 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Recent publications on patient-specific instrumentation for total knee arthroplasty have reported considerable variability in the axial positioning of the cutting guides for the femoral component. These personalized cutting guides are manufactured based on bone shape data, generated from magnetic resonance imaging (MRI) or computed tomography (CT). The goal of this study was to compare the reproducibility and accuracy of distal femoral torsion (DFT) values measured using these two imaging modalities. We hypothesized that MRI does not reproducibly and consistently measure DFT and is not as accurate as CT scan.

Methods

Anonymized radiology records from 54 patients that included MRI and CT scans of the knee were read in random order by two observers, on two separate occasions. These records were from patients being considered for a meniscal or osteochondral graft and who had their knee explored, but who had not undergone femoral or tibial surgery and were free of osteoarthritis. The DFT was estimated using the posterior condylar angle (PCA), using both its anatomical and surgical definitions. The intra- and inter-observer reproducibility of the MRI and the differences relative to CT scan measurements were analysed.

Results

The average intra-observer difference for the MRI evaluation of the anatomical PCA was 0.8±1.2°; it was 0.4±0.9° for the surgical PCA. More than 1° difference from the average was found in 8 cases (14%) using the anatomical PCA measurement and 4 cases (7.4%) when using the surgical PCA (P=0.4). The intraclass correlation coefficients (ICCs) were 0.67 (95% CI: 0.33–0.85) and 0.74 (95% CI: 0.47–0.89) for the anatomical and surgical PCA, respectively. The average inter-observer difference for the MRI evaluation of the anatomical PCA was 1.6±1.4°; it was 1.5±1.0° for the surgical PCA. More than 1° difference from the average was found in 27 cases (50%) using the anatomical PCA measurement and 22 cases (40%) when using the surgical PCA (P=0.4). The ICCs were 0.31 (95% CI: 0.14–0.65) and 0.48 (95% CI: 0.06–0.75) for the anatomical and surgical PCA, respectively. The average differences between the CT and MRI measurements were 1.4±1.1° (0.2–5°) and 1.1±0.8° (0–3.6°) for the anatomical and surgical PCA, respectively. Greater than 1° difference between CT and MRI was found in 29 records (54%) for the anatomical PCA and in 18 records (33%) for the surgical PCA (P=0.03).

Conclusion

DFT measurement on MRI is more reproducible and consistent when using the surgical PCA. MRI measurements differed by more than 1° relative to CT measurements in more than one-third of cases.

Case control study

Level III.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Distal femoral torsion, MRI, CT scan, Patient-specific instrumentation, Total knee arthroplasty


Mappa


© 2015  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 101 - N° 8

P. 937-940 - dicembre 2015 Ritorno al numero
Articolo precedente Articolo precedente
  • Small diameter metal-on-metal total hip arthroplasty at 13 years – a follow-up study
  • N. Tardy, A. Maqdes, P. Boisrenoult, P. Beaufils, P. Oger
| Articolo seguente Articolo seguente
  • Cruciate-sacrificing total knee arthroplasty and insert design: A radiologic study of sagittal laxity
  • B. Appy Fedida, E. Krief, E. Havet, P. Massin, P. Mertl

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.