Ultra-low-dose chest CT with iterative reconstruction does not alter anatomical image quality - 11/05/17

Doi : 10.1016/j.diii.2016.06.009 
F. Macri a, b, c, , J. Greffier a, b, F.R. Pereira a, b, C. Mandoul a, b, E. Khasanova a, b, G. Gualdi c, J.P. Beregi a, b
a Department of radiology, university hospital of Nîmes, 4, rue du Professeur Robert-Debré, 30029 Nîmes, France 
b Research team EA 2415, 4, rue du Professeur Robert-Debré, 30029 Nîmes, France 
c Department of emergency radiology, “Sapienza” university-hospital Um-berto I, 155, viale del Policlinico, 00161 Rome, Italy 

Corresponding author. Department of radiology, university hospital of Nîmes, 4, rue du Professeur Robert-Debré, 30029 Nîmes, France.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Purpose

To evaluate the effect of dose reduction with iterative reconstruction (IR) on image quality of chest CT scan.

Materials and methods

Eighteen human cadavers had chest CT with one reference CT protocol (RP-CT; 120kVp/200mAs) and two protocols with dose reduction: low-dose-CT (LD-CT; 120kVp/40mAs) and ultra-low-dose CT (ULD-CT; 120kVp/10mAs). Data were reconstructed with filter-back-projection (FBP) for RP-CT and with FBP and IR (sinogram affirmed iterative reconstruction [SAFIRE®]) algorithm for LD-CT and ULD-CT. Volume CT dose index (CTDIvol) were recorded. The signal-to-noise (SNR), contrast-to-noise (CNR) ratios of LD-CT and ULD-CT and quantitative parameters were compared to RP-CT. Two radiologists reviewed the CT examinations assessed independently the quality of anatomical structures and expressed a confidence level using a 2-point scale (50% and 95%).

Results

CTDIvol was 2.69 mGy for LD-CT (−80%; P<0.01) and 0.67 mGy for ULD-CT (−95%; P<0.01) as compared to 13.42 mGy for RP-CT. SNR and CNR were significantly decreased (P<0.01) for LD-CT and ULD-CT, but IR improved these values satisfactorily. No significant differences were observed for quantitative measurements. Radiologists rated excellent/good the RP-CT and LD-CT images, whereas good/fair the ULD-CT images. Confidence level for subjective anatomical analysis was 95% for all protocols.

Conclusions

Dose reduction with a dose lower than 1 mGy, used in conjunction with IR allows performing chest CT examinations that provide a high quality of anatomical structures.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Ultra-low-dose chest CT, Sinogram affirmed iterative reconstruction, Image quality, Dose reduction, Safire®


Mappa


© 2016  Editions françaises de radiologie. Pubblicato da Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 97 - N° 11

P. 1131-1140 - novembre 2016 Ritorno al numero
Articolo precedente Articolo precedente
  • Current practice and access to prostate MR imaging in France
  • R. Renard-Penna, O. Rouvière, P. Puech, C. Borgogno, L. Abbas, C. Roy, M. Claudon, J.-M. Correas, L. Cormier, G. Ploussard, A. Mejean, S. Tezenas-du-Montcel, F. Rozet
| Articolo seguente Articolo seguente
  • Prospective ECG-gated high-pitch dual-source cardiac CT angiography in the diagnosis of congenital cardiovascular abnormalities: Radiation dose and diagnostic efficacy in a pediatric population
  • M. Koplay, O. Kizilca, D. Cimen, M. Sivri, H. Erdogan, O. Guvenc, M. Oc, B. Oran

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.