Sagittal reduction of spinal deformity: Superior versus lateral screw-rod connection - 29/10/21

Doi : 10.1016/j.otsr.2021.102954 
Solène Prost a, Sébastien Pesenti b, Kaissar Farah a, Patrick Tropiano a, Stéphane Fuentes a, Benjamin Blondel a,
a Aix-Marseille Université, APHM, CNRS, ISM, CHU Timone, Unité de chirurgie rachidienne, 264, rue Saint-Pierre, 13005 Marseille, France 
b Service de chirurgie orthopédique pédiatrique, Aix-Marseille Université, APHM, CNRS, ISM, CHU Timone, 264, rue Saint-Pierre, 13005 Marseille, France 

Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Introduction

Spinal malalignment can greatly impact a patient's quality of life. Various sagittal parameters are used as realignment goals; however, about 50% of patients end up being under-corrected postoperatively. To improve the correction, prebent rods are available with a radius of curvature corresponding to the patient's “ideal” sagittal alignment. But no studies have been done on how the radius of curvature changes according to the type of connection between the pedicle screws and rods. The goal of this experimental study was to quantify how much prebent rods flatten based on the method used to connect the screw and rod: top-loading screw vs. dome screw with lateral connector.

Methods

The experiment was done on a material testing system in axial compression on three constructs consisting of two rods secured with top-loading screws and three other constructs consisting of two rods secured with dome screws and lateral connector. The maximum angle of the construct was measured during loading and after removing the load. The primary outcome measure was the mean angle in each construct at each step.

Results

The mean angle of the constructs with top-loading screws when subjected to 500 N load was significantly less than in the constructs with dome screws and lateral connector: 18.6° vs. 24.5° respectively (p<0.0003). The mean angle of the constructs with top-loading screws after removing the load was significantly less than in the constructs with dome screws and lateral connector: 25.7° vs. 32.3° respectively, (p<0.0005).

Conclusion

In vitro, top-loading screws produced significantly greater flattening than dome screws with lateral connector. These findings must be confirmed in vivo. Understanding the behavior of rods as a function of the type of screw connection can be an important factor to minimize the risk of under-correction in the sagittal plane.

Level of evidence

III.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Spine, Pedicle screws, Rods, Sagittal alignment


Mappa


© 2021  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 107 - N° 7

Articolo 102954- novembre 2021 Ritorno al numero
Articolo precedente Articolo precedente
  • Treatment of lumbar canal stenosis in patients with compensated sagittal balance
  • Yann Mohsinaly, Louis Boissiere, Cédric Maillot, Sébastien Pesenti, Jean-Charles Le Huec
| Articolo seguente Articolo seguente
  • Reliability assessment of cervical spine parameters measured on full-body radiographs in asymptomatic subjects and patients with spinal deformity
  • Michel Salameh, Aren Joe Bizdikian, Eddy Saad, Renée Maria Saliby, Rhéa Nacouzi, Nour Khalil, Ismat Ghanem, Gaby Kreichati, Ayman Assi

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2025 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.