Mechanical behavior of screw versus Endobutton for coracoid bone-block fixation - 23/09/20

Doi : 10.1016/j.otsr.2020.03.035 
Vadim Azoulay a, b, Jérôme Briot b, Pierre Mansat a, b, Pascal Swider a, b, Nicolas Bonnevialle a, ⁎, b
a Service d’orthopédie, CHU de Toulouse, place Baylac, 31059 Toulouse cedex 09, France 
b Laboratoire de biomécanique, institut de mécanique des fluides de Toulouse UMR CNRS 5502, Toulouse, France 

⁎Corresponding author.

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.
Articolo gratuito.

Si connetta per beneficiarne

Abstract

Introduction

Arthroscopic coracoid bone-block fixation by Endobutton was developed to avoid the complications associated with screwing. However, few studies have assessed the mechanical characteristics of the two. The aim of the present study was to assess and compare fixation rigidity by screw versus Endobutton. The study hypothesis was that rigidity is lower with Endobutton than with screws.

Material and method

3D print-outs of a glenoid and a coracoid process were obtained from CT scans of a patient showing anterior shoulder instability with significant bone defect. Four types of coracoid fixation were implemented: 1 or 2 4.5mm malleolar screws, and 1 or 2 Endobuttons. Three specimens per assembly were placed on a specific test bench. Lateromedial bone-block compression was exerted at 0.1mm/sec at 3 points: superior, central, inferior. The resultant force and bone-block displacement were recorded.

Results

Mean fixation rigidity with 1 screw, 2 screws, 1 Endobutton and 2 Endobuttons was respectively 158N/mm (range, 133–179), 249N/mm (241–259), 10N/mm (5–13) and 14N/mm (13–15), with significant difference between the screw and Endobutton groups (p<0.001). Displacement was greater with 1 than 2 Endobuttons under superior or inferior force, while the difference was non-significant under central force (7.45 vs 6.93mm; p=0.53)

Conclusions

Screw fixation showed greater rigidity, while the Endobutton assembly showed less tension, leading to greater bone-block mobilization. The interest of using two Endobuttons is to reduce displacement under polar pressure. the present biomechanical study confirmed the mechanical vulnerability of bone-blocks fixed by endobutton until consolidation is achieved.

Level of evidence

Biomechanical study.

Il testo completo di questo articolo è disponibile in PDF.

Keywords : Latarjet, Double-button, Endobutton, Bone block, Instability, Shoulder, Biomechanics


Mappa


© 2020  Elsevier Masson SAS. Tutti i diritti riservati.
Aggiungere alla mia biblioteca Togliere dalla mia biblioteca Stampare
Esportazione

    Citazioni Export

  • File

  • Contenuto

Vol 106 - N° 6

P. 1089-1093 - ottobre 2020 Ritorno al numero
Articolo precedente Articolo precedente
  • Basic morphological characteristics of coracoid grafts obtained by open and arthroscopic Latarjet techniques: A comparative study
  • Alvaro Minuesa-Asensio, Francisco García-Esteo, José Ramón Mérida-Velasco, Carmen Barrio-Asensio, Crótida de la Cuadra-Blanco, Jorge Murillo-González
| Articolo seguente Articolo seguente
  • C5/C6 brachial plexus palsy reconstruction using nerve surgery: long-term functional outcomes
  • Laure Pages, Malo Le Hanneur, Adeline Cambon-Binder, Zoubir Belkheyar

Benvenuto su EM|consulte, il riferimento dei professionisti della salute.

@@150455@@ Voir plus

Il mio account


Dichiarazione CNIL

EM-CONSULTE.COM è registrato presso la CNIL, dichiarazione n. 1286925.

Ai sensi della legge n. 78-17 del 6 gennaio 1978 sull'informatica, sui file e sulle libertà, Lei puo' esercitare i diritti di opposizione (art.26 della legge), di accesso (art.34 a 38 Legge), e di rettifica (art.36 della legge) per i dati che La riguardano. Lei puo' cosi chiedere che siano rettificati, compeltati, chiariti, aggiornati o cancellati i suoi dati personali inesati, incompleti, equivoci, obsoleti o la cui raccolta o di uso o di conservazione sono vietati.
Le informazioni relative ai visitatori del nostro sito, compresa la loro identità, sono confidenziali.
Il responsabile del sito si impegna sull'onore a rispettare le condizioni legali di confidenzialità applicabili in Francia e a non divulgare tali informazioni a terzi.


Tutto il contenuto di questo sito: Copyright © 2026 Elsevier, i suoi licenziatari e contributori. Tutti i diritti sono riservati. Inclusi diritti per estrazione di testo e di dati, addestramento dell’intelligenza artificiale, e tecnologie simili. Per tutto il contenuto ‘open access’ sono applicati i termini della licenza Creative Commons.